In vitro and in vivo flow characteristics of glaucoma drainage implants.

J A Prata1, A Mérmoud, L LaBree, D S Minckler.   

Abstract

PURPOSE: To determine pressure-flow characteristics at physiologic flow rates in vitro and in vivo in rabbits for Ahmed, Baerveldt, Krupin disk, and OptiMed glaucoma implants. The Molteno dual-chamber implant also was evaluated in vivo only.
METHODS: Five samples of each glaucoma implant were studied. Baerveldt implants were ligated partially for in vitro testing. Opening and closing pressures in air or after immersion in balanced salt solution or plasma were evaluated for the valved devices (Ahmed and Krupin). Pressures were measured in vitro and in vivo in normal rabbits at flow rates preset at between 2 and 25 microliters/minute after the tubes were connected to a closed manometric system. In vivo measurements were made 24 hours after implantation. Resistance to flow was calculated using Poiseuille's equation after at least three separate flow rate readings.
RESULTS: In air, the Ahmed and Krupin valves had opening pressures of 9.2 +/- 3.4 and 7.2 +/- 0.6 mmHg and closing pressures of 5.2 +/- 0.9 and 3.9 +/- 1 mmHg, respectively. Neither opening nor closing pressures could be determined when Ahmed and Krupin valves were immersed. In vitro, the Ahmed and OptiMed devices had higher pressures than did other devices at a 2-microliters/minute flow rate of balanced salt solution. During perfusion with plasma, only the OptiMed device maintained higher pressures than with balanced salt. With all devices, pressures fell rapidly to zero after flow was stopped. The OptiMed device demonstrated the highest resistance values. In vivo, the Ahmed device provided pressures of 7.5 +/- 0.8 mmHg and the OptiMed device gave pressures of 19.6 +/- 5.6 mmHg at a 2-microliters/minute flow rate. After 15 minutes of flow shutdown, the OptiMed implant maintained pressures of 7.1 +/- 1.1 mmHg. The Baerveldt (nonligatured), Krupin, and Molteno dual-chamber implants had similar resistances and pressures in vivo. Pressures with all devices in vivo fell rapidly to zero after conjunctival wound disruption.
CONCLUSION: Neither the Ahmed nor Krupin devices had demonstrable opening or closing pressures when tested in vitro immersed in balanced salt solution or plasma. With all devices, pressures were higher in vivo than in vitro due to tissue-induced resistance around the explant. Both Ahmed and Krupin valves functioned as flow-restricting devices at the flow rates studied, but did not close after initial perfusion with fluid.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7777296     DOI: 10.1016/s0161-6420(95)30937-2

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  27 in total

1.  Survival analysis for success of Molteno tube implants.

Authors:  D C Broadway; M Iester; M Schulzer; G R Douglas
Journal:  Br J Ophthalmol       Date:  2001-06       Impact factor: 4.638

2.  Baerveldt implant in refractory glaucoma: long-term results and factors influencing outcome.

Authors:  S Roy; E Ravinet; A Mermoud
Journal:  Int Ophthalmol       Date:  2001       Impact factor: 2.031

Review 3.  Aqueous shunts for glaucoma.

Authors:  D S Minckler; S S Vedula; T J Li; M C Mathew; R S Ayyala; B A Francis
Journal:  Cochrane Database Syst Rev       Date:  2006-04-19

4.  Clinical experience of e-PTFE membrane implant surgery for refractory glaucoma.

Authors:  C Kim; Y Kim; S Choi; S Lee; B Ahn
Journal:  Br J Ophthalmol       Date:  2003-01       Impact factor: 4.638

Review 5.  Glaucoma drainage devices; past, present, and future.

Authors:  K S Lim; B D Allan; A W Lloyd; A Muir; P T Khaw
Journal:  Br J Ophthalmol       Date:  1998-09       Impact factor: 4.638

Review 6.  Aqueous shunts for glaucoma.

Authors:  Victoria L Tseng; Anne L Coleman; Melinda Y Chang; Joseph Caprioli
Journal:  Cochrane Database Syst Rev       Date:  2017-07-28

7.  Efficacy and safety of the Ahmed glaucoma valve implant in Chinese eyes with complicated glaucoma.

Authors:  J S Lai; A S Poon; J K Chua; C C Tham; A T Leung; D S Lam
Journal:  Br J Ophthalmol       Date:  2000-07       Impact factor: 4.638

8.  Cell and protein adhesion studies in glaucoma drainage device development. The AGFID project team.

Authors:  K S Lim
Journal:  Br J Ophthalmol       Date:  1999-10       Impact factor: 4.638

9.  Assessing risk factors for postoperative hypotony in Ahmed glaucoma valve implantation surgery.

Authors:  Ahmet Kaderli; Gülizar Demirok; Güner Üney; Mehmet Yakın; Berku Günal; Ümit Ekşioğlu
Journal:  Int Ophthalmol       Date:  2021-05-21       Impact factor: 2.031

10.  Risk factors for flat anterior chamber after glaucoma filtration surgery.

Authors:  Xiang-Ji Li; Richard Filek; Xiang-Ge He; Wei Wang; Hong Liu; Lian He; Yu-Rong Tang; Lin Xie
Journal:  Int J Ophthalmol       Date:  2018-08-18       Impact factor: 1.779

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.