Literature DB >> 7767320

An automated [15Q]H2O production and injection system for PET imaging.

B M Palmer1, M Sajjad, D A Rottenberg.   

Abstract

A computer controlled [15O]H2O injection system has been designed for multiple bolus injections during PET cerebral blood flow studies. The system is housed in a lead safe, and radiation exposure to nearby personnel is low; we have measured rates of approx. 100 microSv/h (10 mR/h) during 2 min of [15O]H2O production and < 50 microSv/h (< 5 mR/h) after 3.7 GBq (100 mCi) has been produced. Reproducibility studies show that the injected activity varies with a coefficient of variation of 2-3%.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7767320     DOI: 10.1016/0969-8051(94)00085-x

Source DB:  PubMed          Journal:  Nucl Med Biol        ISSN: 0969-8051            Impact factor:   2.408