Evaluation of a transit-time system for the chronic measurement of blood flow in conscious sheep.

J A Bednarik1, C N May.   

Abstract

The accuracy of transit-time ultrasonic flow probes for measurement of regional blood flow and cardiac output was evaluated after long-term implantation in sheep. Transit-time flow probes (3, 4, 6, and 20 mm) accurately measured flow in vitro. Recalibration in vivo demonstrated that this accuracy was maintained after 1-9 mo of implantation on the left circumflex coronary (3-mm probe), cranial mesenteric (6-mm probe), left renal (4-mm probe), and left external iliac (6-mm probe) arteries of sheep. The flow probes also showed good zero stability. However, a transit-time flow probe (20 mm) chronically implanted on the pulmonary trunk significantly underestimated cardiac output compared with thermodilution or timed collection of blood. Although this flow probe underestimated flow, the response was linear. Bilateral carotid occlusion caused mesenteric, renal, and iliac vasoconstrictions, confirming that innervation of these vascular beds was undamaged. For experimental purposes, regional blood flow was measured with transit-time flow probes and cardiac output was measured with electromagnetic flow probes calibrated against thermodilution. In summary, transit-time flow probes reliably and accurately measure regional blood flow over many months in adult sheep, but, to measure cardiac output in sheep, the probes must be calibrated in vivo against another reference technique.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7759421     DOI: 10.1152/jappl.1995.78.2.524

Source DB:  PubMed          Journal:  J Appl Physiol (1985)        ISSN: 0161-7567


  9 in total

1.  An assessment of the accuracy of renal blood flow estimation by Doppler ultrasound.

Authors:  Li Wan; Natalie Yang; Chee-Yan Hiew; Anthony Schelleman; Lynne Johnson; Clive May; Rinaldo Bellomo
Journal:  Intensive Care Med       Date:  2008-04-12       Impact factor: 17.440

2.  The haemodynamic and metabolic effects of epinephrine in experimental hyperdynamic septic shock.

Authors:  David Di Giantomasso; Rinaldo Bellomo; Clive N May
Journal:  Intensive Care Med       Date:  2005-02-15       Impact factor: 17.440

3.  Real-time measurement of renal blood flow in healthy subjects using contrast-enhanced ultrasound.

Authors:  Kambiz Kalantarinia; J Todd Belcik; James T Patrie; Kevin Wei
Journal:  Am J Physiol Renal Physiol       Date:  2009-07-22

4.  The effects of increasing doses of noradrenaline on systemic and renal circulations in acute bacteraemic dogs.

Authors:  Zhi-Yong Peng; Lester A H Critchley; Benny S P Fok
Journal:  Intensive Care Med       Date:  2005-08-09       Impact factor: 17.440

5.  Major influence of a 'smoke and mirrors' effect caused by wave reflection on early diastolic coronary arterial wave intensity.

Authors:  Jonathan P Mynard; Daniel J Penny; Joseph J Smolich
Journal:  J Physiol       Date:  2018-02-13       Impact factor: 5.182

6.  The effect of normal saline resuscitation on vital organ blood flow in septic sheep.

Authors:  Li Wan; Rinaldo Bellomo; Clive N May
Journal:  Intensive Care Med       Date:  2006-06-15       Impact factor: 17.440

7.  Norepinephrine and vital organ blood flow during experimental hyperdynamic sepsis.

Authors:  David Di Giantomasso; Clive N May; Rinaldo Bellomo
Journal:  Intensive Care Med       Date:  2003-04-16       Impact factor: 17.440

8.  Direct coronary vasodilator action of adrenomedullin is mediated by nitric oxide.

Authors:  R De Matteo; C N May
Journal:  Br J Pharmacol       Date:  2003-11-17       Impact factor: 8.739

9.  Angiotensin II in experimental hyperdynamic sepsis.

Authors:  Li Wan; Christoph Langenberg; Rinaldo Bellomo; Clive N May
Journal:  Crit Care       Date:  2009-11-30       Impact factor: 9.097

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.