K Matsui1, Y Oda, H Ohe, S Tanaka, N Asakawa.
Abstract
High-performance liquid chromatography with column switching and mass spectrometry (MS) was applied to the on-line determination and resolution of the enantiomers of E2020 (acetylcholinesterase inhibitor) in plasma. This system employs two avidin columns and fast atom bombardment (FAB)-MS. A plasma sample was injected directly into an avidin trapping column (10 mm x 4.0 mm I.D.). The plasma protein was washed out from the trapping column immediately while E2020 was retained. After the column-switching procedure, E2020 was separated enantioselectivity in an avidin analytical column. The separated E2020 enantiomers were specifically detected by FAB-MS without interference from metabolites of E2020 and plasma constituents. The limit of quantification for each enantiomer of E2020 in plasma was 1.0 ng/ml and the intra- and inter-assay relative standard deviations for the method were less than 5.2%. The assay was validated for enantioselective pharmacokinetic studies in the dog.Entities:
Mesh:
Substances:
Year: 1995 PMID: 7719466 DOI: 10.1016/0021-9673(94)01005-y
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759