| Literature DB >> 7710970 |
Abstract
A retrospective study of fractures of peritoneal catheters of ventriculoperitoneal shunts is described. Fracture of the peritoneal catheter accounts for 21.2% of shunt revisions. During the time period of January 1985 through July 1993, 401 PS Medical peritoneal catheters were installed and 23 fractures occurred. The fracture rate of PS Medical small-sized catheters is 15.3%. The fracture rate of PS Medical regular-sized catheters is 2.1%. The overall survival rates of the PS Medical small-sized and regular-sized catheters are significantly statistically different (p < 0.001). The peritoneal catheters fractured most frequently in the neck.Entities:
Mesh:
Year: 1995 PMID: 7710970 DOI: 10.1159/000120884
Source DB: PubMed Journal: Pediatr Neurosurg ISSN: 1016-2291 Impact factor: 1.162