Calibrating and testing tissue equivalent proportional counters with 37Ar.

E Anachkova1, A M Kellerer, H Roos.   

Abstract

A method for testing and calibrating tissue equivalent proportional counters with 37Ar is described. 37Ar is produced by exposure of argon in its normal isotope composition to thermal neutrons. It is shown that--up to volume ratios of 0.01 of argon to the tissue equivalent gas--there is no appreciable effect of the argon admixture on the function of the proportional counter. Conventional calibration methods with characteristic x-rays or with alpha-particles require modifications of the detectors, and they test only small sub-volumes in the counters. In contrast, argon permits calibrations and tests of the resolution that are representative for the entire counter volume and that do not require changes in detector construction. The method is equally applicable to multi-element proportional counters; it is here exemplified by its application to a long cylindrical counter of simplified design that is part of such a multi-element configuration.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7708908     DOI: 10.1007/bf01210456

Source DB:  PubMed          Journal:  Radiat Environ Biophys        ISSN: 0301-634X            Impact factor:   1.925


  4 in total

1.  W values and other transport data on low energy electrons in tissue equivalent gas.

Authors:  E Waibel; B Grosswendt
Journal:  Phys Med Biol       Date:  1992-05       Impact factor: 3.609

2.  Calculation of photon mass energy-transfer and mass energy-absorption coefficients.

Authors:  S M Seltzer
Journal:  Radiat Res       Date:  1993-11       Impact factor: 2.841

3.  Photon mass attenuation and mass energy-absorption coefficients for H, C, N, O, Ar, and seven mixtures from 0.1 keV to 20 MeV.

Authors:  J H Hubbell
Journal:  Radiat Res       Date:  1977-04       Impact factor: 2.841

4.  Multi-element dosimeters for radiation protection measurements.

Authors:  H H Rossi
Journal:  Health Phys       Date:  1983-04       Impact factor: 1.316

  4 in total
  1 in total

1.  Effect of wire eccentricity on the performance of cylindrical proportional counters.

Authors:  E Anachkova; H Roos; E Nekolla; A M Kellerer
Journal:  Radiat Environ Biophys       Date:  1995-11       Impact factor: 1.925

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.