Evaluation of temporary atrial pacing leads.

W A Scott1, M Lemler, L Farley, R Zehr.   

Abstract

Thirty-five patients were randomized to receive either the Medtronic 6500 or one of two braided multifilament temporary pacing leads in the atrium following open heart surgery. Sensing performance was judged by amplitude, slew rate, and the proportion of patients with an adequate sensing threshold. Pacing performance was assessed with measures of impedance, threshold voltage, current, and energy. The Medtronic 6500 demonstrated superior sensing and lower energy consumption compared to braided multifilament leads. This type of lead may offer advantages when using atrial synchronous temporary pacing systems.

Entities:  

Mesh:

Year:  1993        PMID: 7692410     DOI: 10.1111/j.1540-8159.1993.tb01812.x