Literature DB >> 7681575

Complications of implantable cardioverter defibrillator therapy: follow-up of 241 patients.

W Grimm1, B F Flores, F E Marchlinski.   

Abstract

In order to determine the incidence of complications of implantable cardioverter defibrillator (ICD) therapy, 241 patients with a total of 353 ICD implantations were followed for 24 +/- 20 months. Complications were defined as any untoward effects experienced by the patient related to the ICD implantation and function or death within 4 weeks of implant or before hospital discharge. During follow-up, 129 of 241 patients (53%) had a total of 166 complications. An operative procedure was required to correct the complication in 50 of the 241 patients (21%). No patient died intraoperatively, but eight patients died within 4 weeks postoperatively or before hospital discharge. ICD infection required removal of the device in 13 patients (5%). Twenty-six patients (11%) had postoperative respiratory complications. Postoperative bleeding and/or thrombosis occurred in 11 patients (4%). Endocardial lead migration was observed in nine patients (4%) and lead adapter or insulation break occurred in nine patients (4%). Fifty-four patients (22%) experienced ECG-documented discharges for non-VT rhythm. In conclusion, although perioperative mortality in ICD patients is low, the majority of patients have complications during follow-up. Recognition of these complications may allow for advances in ICD technology and management strategies to avoid their occurrence.

Entities:  

Mesh:

Year:  1993        PMID: 7681575     DOI: 10.1111/j.1540-8159.1993.tb01565.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  12 in total

1.  Prospective evaluation of the quality and long-term stability of atrial signals in non-thoracotomy defibrillation electrodes: comparison of four different endocardial electrograms.

Authors:  C Wolpert; W Jung; S Spehl; B Schumacher; H Omran; B Esmailzadeh; B Lüderitz
Journal:  J Interv Card Electrophysiol       Date:  1998-12       Impact factor: 1.900

2.  Subcutaneous implantable cardioverter-defibrillator implantation in a patient with a left ventricular assist device already in place.

Authors:  Ashwani Gupta; Faiz Subzposh; Shelley R Hankins; Steven P Kutalek
Journal:  Tex Heart Inst J       Date:  2015-04-01

3.  Outcome of cardioverter-defibrillator implant in patients with arrhythmogenic right ventricular cardiomyopathy.

Authors:  Giuseppe Boriani; Paolo Artale; Mauro Biffi; Cristian Martignani; Lorenzo Frabetti; Cinzia Valzania; Igor Diemberger; Matteo Ziacchi; Matteo Bertini; Claudio Rapezzi; Mario Parlapiano; Angelo Branzi
Journal:  Heart Vessels       Date:  2007-05-21       Impact factor: 2.037

4.  Experience with unipolar pectoral defibrillation.

Authors:  R K Reddy; G H Bardy
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

5.  Inappropriate shocks in a patient treated with a cardioverter defibrillator.

Authors:  W Hutse; R Tavernier; L Jordaens
Journal:  Heart       Date:  1997-04       Impact factor: 5.994

Review 6.  Subcutaneous chronic implantable defibrillation systems in humans.

Authors:  Riccardo Cappato; Warren M Smith; Margaret A Hood; Ian G Crozier; Luc Jordaens; Stefan G Spitzer; Andrey V Ardashev; Lucas Boersma; Pierpaolo Lupo; Andrew A Grace; Gust H Bardy
Journal:  J Interv Card Electrophysiol       Date:  2012-03-17       Impact factor: 1.900

7.  Implantable cardioverter-defibrillators in patients with arrhythmogenic right ventricular cardiomyopathy: the course of electronic parameters, clinical features, and complications during long-term follow-up.

Authors:  Giacomo Mugnai; Ruggero Tomei; Clementina Dugo; Luca Tomasi; Giovanni Morani; Corrado Vassanelli
Journal:  J Interv Card Electrophysiol       Date:  2014-06-14       Impact factor: 1.900

8.  Age dependent efficacy of implantable cardioverter-defibrillator treatment: observations in 450 patients over an 11 year period.

Authors:  H J Trappe; P Pfitzner; M Achtelik; H G Fieguth
Journal:  Heart       Date:  1997-10       Impact factor: 5.994

9.  [Influence of waveform and configuration of electrodes on the defibrillation threshold of implantable cardioverter-defibrillators].

Authors:  M Block; D Hammel; G Breithardt
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03

10.  Benefits of treatment with implantable cardioverter-defibrillators in patients with stable ventricular tachycardia without cardiac arrest.

Authors:  D Böcker; M Block; F Isbruch; C Fastenrath; M Castrucci; D Hammel; H H Scheld; M Borggrefe; G Breithardt
Journal:  Br Heart J       Date:  1995-02
View more