Anatomical and morphological evaluation of pacemaker lead compression.

D M Jacobs1, A S Fink, R P Miller, W R Anderson, R D McVenes, J F Lessar, K E Cobian, D B Staffanson, J E Upton, M P Bubrick.   

Abstract

In recent years, pacemaker lead failure due to compressive damage has been reported with increasing frequency. To document the mechanism of this failure, we evaluated explanted mechanically damaged leads with electrical testing, optical microscopy, and in some cases, scanning electron microscopy (SEM). In addition, we performed an autopsy study to measure the compressive loads on catheters placed percutaneously through the costoclavicular angle, as well as by cephalic cutdown. Of the 49 explanted compression damaged leads with enough clinical data for analysis, all had been placed by percutaneous subclavian puncture. Our autopsy data confirmed the significant increase in pressures generated in the costoclavicular angle for medial percutaneous subclavian catheterization (126 +/- 26 mmHg) compared to a more lateral percutaneous subclavian puncture (63 +/- 15 mmHg) or a cephalic cutdown (38 +/- 13 mmHg) (P < 0.01). In vivo coil compression testing documented loads up to 100 pounds per linear inch of coil and a compressive morphology by SEM identical to that seen in the clinical explants. Pacemaker leads appear to be susceptible to compression damage when placed by subclavian venipuncture. When possible, leads should be placed such that they avoid the tight costoclavicular angle.

Entities:  

Mesh:

Year:  1993        PMID: 7681195     DOI: 10.1111/j.1540-8159.1993.tb01606.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  6 in total

1.  Pacemaker lead entrapment syndrome.

Authors:  Kamal Gupta; Rollo P Villareal; Abdi Rasekh; Ali Massumi
Journal:  Tex Heart Inst J       Date:  2003

Review 2.  Radiology of cardiac devices and their complications.

Authors:  J Dipoce; A Bernheim; H Spindola-Franco
Journal:  Br J Radiol       Date:  2014-11-20       Impact factor: 3.039

3.  Subpectoral implantation of a cardioverter defibrillator under local anaesthesia.

Authors:  K J Lipscomb; N J Linker; A P Fitzpatrick
Journal:  Heart       Date:  1998-03       Impact factor: 5.994

4.  Prospective study to develop surface landmarks for blind axillary vein puncture for permanent pacemaker and defibrillator lead implantation and compare it to available contrast venography guided technique.

Authors:  S Mehrotra; Manoj Kumar Rohit
Journal:  Indian Heart J       Date:  2015-05-13

Review 5.  Sport Participation in Patients with Implantable Cardioverter-Defibrillators.

Authors:  Rachel Lampert
Journal:  Curr Treat Options Cardiovasc Med       Date:  2019-11-14

6.  Possible complications of subclavian crush syndrome.

Authors:  S A M Said; C H J M Ticheler; C M Stassen; A Derks; H T Droste
Journal:  Neth Heart J       Date:  2005-03       Impact factor: 2.380

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.