Hepatic enhancement during helical CT: a comparison of moderate rate uniphasic and biphasic contrast injection protocols.

B A Birnbaum1, J E Jacobs, D Yin.   

Abstract

OBJECTIVE: The purpose of this study was to compare the degree and timing of peak hepatic enhancement, optimal scanning intervals, and optimal delay times of moderate-rate uniphasic and biphasic contrast material injection protocols for hepatic helical CT.
MATERIALS AND METHODS: One hundred fifty patients were randomized into three injection protocols, receiving 42.3 g iodine (150 ml iothalamate meglumine) delivered using 3 ml/sec uniphasic, 2 ml/sec uniphasic, or biphasic (3 ml/sec [50 ml], 1 ml/sec [100 ml]) technique. Statistically fitted aortic and hepatic enhancement curves were generated from dynamic incremental CT data for each patient. Protocols were compared by maximum hepatic enhancement, and contrast enhancement indices were modeled for a 38-sec helical acquisition.
RESULTS: The 3 ml/sec and 2 ml/sec uniphasic protocols produced higher peak hepatic enhancement (64 +/- 15 H and 62 +/- 15 H [mean +/- 1 SD]) than the 3 ml/sec biphasic protocol (52 +/- 10 H; p < .001). Contrast enhancement indices for the 3 ml/sec uniphasic and 2 ml/sec uniphasic protocols (385 +/- 398 H/sec and 397 +/- 412 H/sec) were significantly greater than the index for the 3 ml/sec biphasic protocol (123 +/- 194 H/sec; p < .0001) at a 50-H threshold. Optimal scan delay times were 50 +/- 8, 75 +/- 7, and 119 +/- 8 sec, respectively, for the 3 ml/sec uniphasic, 2 ml/sec uniphasic, and 3 ml/sec biphasic techniques.
CONCLUSION: The moderate-rate uniphasic injections studied provided greater hepatic enhancement throughout the helical acquisition without requiring the prohibitively long delay time necessitated by the moderate-rate biphasic injection. These findings differ from prior results that showed that a uniphasic injection may provide comparable levels of hepatic enhancement when compared with a high-flow-rate biphasic injection.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7676980     DOI: 10.2214/ajr.165.4.7676980

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  2 in total

1.  CT angiography covering both cervical and cerebral arteries using high iodine concentration contrast material with dose reduction on a 16 multidetector-row system.

Authors:  Akira Fujikawa; Kazuhiro Tsuchiya; Masamichi Imai; Toshiaki Nitatori
Journal:  Neuroradiology       Date:  2009-10-16       Impact factor: 2.804

2.  Contrast media in abdominal computed tomography: optimization of delivery methods.

Authors:  J K Han; B I Choi; A Y Kim; S J Kim
Journal:  Korean J Radiol       Date:  2001 Jan-Mar       Impact factor: 3.500

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.