Thermal response and torque resistance of five cortical half-pins under simulated insertion technique.

M A Wikenheiser1, M D Markel, D G Lewallen, E Y Chao.   

Abstract

A model was developed that can quantitate heat generation during placement of half-pins in cortical bone. Five half-pins were tested to assess differences in insertion torque, heat generation, and micro-damage at the pin-bone interface. Thin thermocouple probes were placed 0.5 mm from the track of the pin and within the pin to measure its temperature during insertion. Scanning electron microscopy was used to view the pin-bone interface to assess the microdamage during placement. The design of the tip of the pin influenced insertion torque and heat generation. Higher heat generation was measured when a thermocouple was placed within the pin itself and less was measured when thermocouple probes were placed within bone samples 0.5 mm from the impending pin track. Furthermore, insertion torque and thermal responses were related, but there were no significant differences in microdamage to bone when different pins and drilling/tapping techniques were used. Due to the significant heat generation at the pin-bone interface, proper cooling with saline irrigation should be applied during pin insertion regardless of the design of the pin. The microdamage observed at the surface of the pin track may have significant implications with regard to loosening of pins, but such effects must be studied with in vivo models.

Entities:  

Mesh:

Year:  1995        PMID: 7674078     DOI: 10.1002/jor.1100130418

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  9 in total

1.  Pearls and pitfalls of deformity correction and limb lengthening via monolateral external fixation.

Authors:  K J Noonan; C T Price
Journal:  Iowa Orthop J       Date:  1996

2.  Effect of the starting point of half-pin insertion on the insertional torque of the pin at the tibia.

Authors:  Sung Jae Kim; Sung Hwan Kim; Young Hwan Kim; Yong Min Chun
Journal:  Yonsei Med J       Date:  2015-01       Impact factor: 2.759

Review 3.  Prevention of pin site infection in external fixation: a review of the literature.

Authors:  Nikolas H Kazmers; Austin T Fragomen; S Robert Rozbruch
Journal:  Strategies Trauma Limb Reconstr       Date:  2016-05-12

4.  Improving tribological and anti-bacterial properties of titanium external fixation pins through surface ceramic conversion.

Authors:  Huan Dong; Tatiana Mukinay; Maojun Li; Richard Hood; Sein Leung Soo; Simon Cockshott; Rachel Sammons; Xiaoying Li
Journal:  J Mater Sci Mater Med       Date:  2016-11-24       Impact factor: 3.896

Review 5.  Biomechanical analysis of four external fixation pin insertion techniques.

Authors:  Dillon Arango; Nathan Tiedeken; Benjamin Clippinger; Solomon Praveen Samuel; Vilas Saldanha; Gene Shaffer
Journal:  Orthop Rev (Pavia)       Date:  2017-10-03

6.  Intraosseous Temperature Change during Installation of Dental Implants with Two Different Surfaces and Different Drilling Protocols: An In Vivo Study in Sheep.

Authors:  Michele Stocchero; Yohei Jinno; Marco Toia; Marianne Ahmad; Evaggelia Papia; Satoshi Yamaguchi; Jonas P Becktor
Journal:  J Clin Med       Date:  2019-08-11       Impact factor: 4.241

7.  Marginal bone resorption of posterior mandible dental implants with different insertion methods.

Authors:  Ehsan Aliabadi; Saeid Tavanafar; Mohammad Saleh Khaghaninejad
Journal:  BMC Oral Health       Date:  2020-01-31       Impact factor: 2.757

8.  Unicortical self-drilling external fixator pins reduce thermal effects during pin insertion.

Authors:  Markus Greinwald; Patrick A Varady; Peter Augat
Journal:  Eur J Trauma Emerg Surg       Date:  2017-12-14       Impact factor: 3.693

9.  Improving the Tribological Properties and Biocompatibility of Zr-Based Bulk Metallic Glass for Potential Biomedical Applications.

Authors:  Victoria Sawyer; Xiao Tao; Huan Dong; Behnam Dashtbozorg; Xiaoying Li; Rachel Sammons; Han-Shan Dong
Journal:  Materials (Basel)       Date:  2020-04-22       Impact factor: 3.623

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.