Evaluation of dose variation during total skin electron irradiation using thermoluminescent dosimeters.

R D Weaver1, B J Gerbi, K E Dusenbery.   

Abstract

PURPOSE: To determine acceptable dose variation using thermoluminescent dosimeters (TLD) in the treatment of Mycosis Fungoides with total skin electron beam (TSEB) irradiation. METHODS AND MATERIALS: From 1983 to 1993, 22 patients were treated with total skin electron beam therapy in the standing position. A six-field technique was used to deliver 2 Gy in two days, treating 4 days per week, to a total dose of 35 to 40 Gy using a degraded 9 MeV electron beam. Thermoluminescent dosimeters were placed on several locations of the body and the results recorded. The variations in these readings were analyzed to determine normal dose variation for various body locations during TSEB.
RESULTS: The dose to flat surfaces of the body was essentially the same as the dose to the prescription point. The dose to tangential surfaces was within +/- 10% of the prescription dose, but the readings showed much more variation (up to 24%). Thin areas of the body showed large deviations from the prescription dose along with a large amount of variation in the readings (up to 22%). Special areas of the body, such as the perineum and eyelid, showed large deviations from the prescription dose with very large (up to 40%) variations in the readings. DISCUSSION: The TLD results of this study will be used as a quality assurance check for all new patients treated with TSEB. The results of the TLDs will be compared with this baseline study to determine if the delivered dose is within acceptable ranges. If the TLD results fall outside the acceptable limits established above, then the patient position can be modified or the technique itself evaluated.

Entities:  

Mesh:

Year:  1995        PMID: 7673036     DOI: 10.1016/0360-3016(95)00161-Q

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  20 in total

1.  Cherenkov imaging for Total Skin Electron Therapy - an evaluation of dose uniformity.

Authors:  Timothy C Zhu; Yihong Ong; Hongjin Sun; Weili Zhong; Tianshun Miao; Andreea Dimofte; Petr Bruza; Amit Maity; John P Plastaras; Ima Paydar; Lei Dong; Brian W Pogue
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-03-30

2.  Monte Carlo (MC) study of dose distribution and Cherenkov imaging in total skin electron therapy (TSET) with TOPAS.

Authors:  Weili Zhong; Yi Hong Ong; Timothy Zhu
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-03-12

Review 3.  Review of the results of the in vivo dosimetry during total skin electron beam therapy.

Authors:  Gabriele Guidi; Giovanni Gottardi; Paola Ceroni; Tiziana Costi
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-15

4.  Dose optimization of total or partial skin electron irradiation by thermoluminescent dosimetry.

Authors:  Lars Schüttrumpf; Klement Neumaier; Cornelius Maihoefer; Maximilian Niyazi; Ute Ganswindt; Minglun Li; Peter Lang; Michael Reiner; Claus Belka; Stefanie Corradini
Journal:  Strahlenther Onkol       Date:  2018-01-19       Impact factor: 3.621

5.  Camera selection for real-time in vivo radiation treatment verification systems using Cherenkov imaging.

Authors:  Jacqueline M Andreozzi; Rongxiao Zhang; Adam K Glaser; Lesley A Jarvis; Brian W Pogue; David J Gladstone
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

6.  Monte Carlo simulation of Cherenkov imaging for Total Skin Electron Treatment with CT DICOM realistic patient geometry.

Authors:  Weili Zhong; Yihong Ong; Tianshu Miao; Brian W Pogue; Timothy C Zhu
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2022-03-04

7.  Cherenkov imaging for total skin electron therapy (TSET).

Authors:  Yunhe Xie; Heather Petroccia; Amit Maity; Tianshun Miao; Yihua Zhu; Petr Bruza; Brian W Pogue; John P Plastaras; Lei Dong; Timothy C Zhu
Journal:  Med Phys       Date:  2019-11-26       Impact factor: 4.071

8.  Flexible optically stimulated luminescence band for 1D in vivo radiation dosimetry.

Authors:  Tae Jin Kim; Kyung Oh Jung; Benjamin Fahimian; Guillem Pratx
Journal:  Phys Med Biol       Date:  2018-08-10       Impact factor: 3.609

9.  Helical irradiation of the total skin with dose painting to replace total skin electron beam therapy for therapy-refractory cutaneous CD4+ T-cell lymphoma.

Authors:  Chen-Hsi Hsieh; Pei-Wei Shueng; Shih-Chiang Lin; Hui-Ju Tien; An-Cheng Shiau; Yueh-Hung Chou; Meng-Hao Wu; Jen-Yu Wang; Chi-Kuan Chen; Yu-Jen Chen
Journal:  Biomed Res Int       Date:  2013-09-23       Impact factor: 3.411

10.  In vivo monitoring of total skin electron dose using optically stimulated luminescence dosimeters.

Authors:  Tanya Kairn; Rachael Wilks; Liting Yu; Craig Lancaster; Scott B Crowe
Journal:  Rep Pract Oncol Radiother       Date:  2019-12-16
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.