Comparison of "look-alike" implant prosthetic retaining screws.

M J Jaarda1, M E Razzoog, D G Gratton.   

Abstract

PURPOSE: The maximum preload torque of implant prosthetic retaining screws from four manufacturers and of two alloy types was measured to determine one index of interchangeability of intersystem components.
MATERIALS AND METHODS: Implant prosthetic retaining screws from four manufacturers (3i Implant Innovations Inc, West Palm Beach, FL; Impla-Med Inc, Sunrise, FL; Nobelpharma USA Inc, Chicago, IL; and Implant Support Systems Inc, Irvine, CA) and of two metal types (gold and titanium) were investigated using an in vitro simulation model. Five screws of each type were tightened down against a gold cylinder using a Tohnichi BTG-6 torque gauge (Tohnichi American Corporation, Northbrook, IL) until fracture occurred.
RESULTS: The 3i Implant Innovations gold and the Nobelpharma gold were not significantly different. The 3i Implant Innovations titanium and the Impla-Med gold were able to withstand less preload torque than the 3i Implant Innovations gold and the Nobelpharma gold. The Implant Support Systems titanium was able to withstand significantly more preload torque than all of the other screws.
CONCLUSIONS: Interchanging implant prosthetic retaining screws could introduce new and unknown variables that may affect the long-term survival of implant fixtures and/or the implant prostheses.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7670607     DOI: 10.1111/j.1532-849x.1995.tb00310.x

Source DB:  PubMed          Journal:  J Prosthodont        ISSN: 1059-941X            Impact factor:   2.752


  1 in total

1.  A Simple and Cost Effective Method used for Removal of a Fractured Implant Abutment Screw: A Case Report.

Authors:  Parth Satwalekar; K Subash Chander; B Anantha Reddy; N Sandeep; N Sandeep; Tanushree Satwalekar
Journal:  J Int Oral Health       Date:  2013-10-26
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.