L C Stephens1, W D Haire, G D Kotulak. 1. Department of Internal Medicine, University of Nebraska Medical Center, Omaha 68198-1210, USA.
Abstract
BACKGROUND: Two hundred dysfunctional central venous catheters used for total parenteral nutrition and administration of cancer chemotherapy were radiographically examined in order to objectively identify thrombotic occlusions as the cause of catheter dysfunction. METHODS: Outcomes of radiographic dye injections were compared with factors such as the inability to aspirate blood or to infuse fluids, catheter type, and duration of catheter placement. RESULTS: Catheter type and duration of placement were not significant factors for predicting the type of dysfunction. Failure to withdraw blood was associated with 96% of the thrombosed catheters; this was also associated with 65% of the catheters with nonthrombotic dysfunctions. Once the cause of catheter occlusion was correctly identified, 90% of the catheters were restored to normal function. CONCLUSIONS: Inability to withdraw blood from a catheter does not necessarily mean it is occluded by thrombus. Mechanical complications account for a significant portion of dysfunctional catheters.
BACKGROUND: Two hundred dysfunctional central venous catheters used for total parenteral nutrition and administration of cancer chemotherapy were radiographically examined in order to objectively identify thrombotic occlusions as the cause of catheter dysfunction. METHODS: Outcomes of radiographic dye injections were compared with factors such as the inability to aspirate blood or to infuse fluids, catheter type, and duration of catheter placement. RESULTS: Catheter type and duration of placement were not significant factors for predicting the type of dysfunction. Failure to withdraw blood was associated with 96% of the thrombosed catheters; this was also associated with 65% of the catheters with nonthrombotic dysfunctions. Once the cause of catheter occlusion was correctly identified, 90% of the catheters were restored to normal function. CONCLUSIONS: Inability to withdraw blood from a catheter does not necessarily mean it is occluded by thrombus. Mechanical complications account for a significant portion of dysfunctional catheters.
Authors: Jacquelyn L Baskin; Ulrike Reiss; Judith A Wilimas; Monika L Metzger; Raul C Ribeiro; Ching-Hon Pui; Scott C Howard Journal: Haematologica Date: 2011-12-16 Impact factor: 9.941
Authors: Jacquelyn L Baskin; Ching-Hon Pui; Ulrike Reiss; Judith A Wilimas; Monika L Metzger; Raul C Ribeiro; Scott C Howard Journal: Lancet Date: 2009-07-11 Impact factor: 79.321
Authors: Asheley B Wathen; Daniel D Myers; Paul Zajkowski; Graham Flory; F Claire Hankenson Journal: J Am Assoc Lab Anim Sci Date: 2009-09 Impact factor: 1.232
Authors: Giovanna Figueiredo; Teresa Fiebig; Stefanie Kirschner; Omid Nikoubashman; Lisa Kabelitz; Ahmed Othman; Andrea Nonn; Martin Kramer; Marc A Brockmann Journal: PLoS One Date: 2015-06-22 Impact factor: 3.240