| Literature DB >> 7655613 |
B Liu1, D Anderson, D R Ferry, L W Seymour, P G de Takats, D J Kerr.
Abstract
A method is reported for the measurement of quercetin in human plasma using reversed-phase high-performance liquid chromatography (HPLC). Quercetin and kaempferol (as internal standard) were spiked into plasma samples and extracted using C18 Sep-Pak Light cartridges (efficiency > 85%). Flavonoids were eluted with aqueous acetone (50% v/v, pH 3.5), dried down and redissolved in aqueous acetone (45% v/v, pH 3.5). The increased osmolarity promoted a phase separation and the water-saturated acetone layer, containing the flavonoids, was analysed by HPLC with aqueous acetone mobile phase (45% v/v acetone in 250 mM sodium dihydrogen sulphate. The mixture was adjusted to pH 3.5 with phosphoric acid and used at a flow-rate of 1.0 ml/min) and mu Bondapak C18 column (150 x 3.9 mm I.D., 10 microns particle size). The detection limit (A375 nm) for quercetin in plasma was 0.1 microgram/ml (300 nM). The method also detects metabolites of quercetin, although these are not yet identified.Entities:
Mesh:
Substances:
Year: 1995 PMID: 7655613 DOI: 10.1016/0378-4347(94)00549-k
Source DB: PubMed Journal: J Chromatogr B Biomed Appl ISSN: 1572-6495