Literature DB >> 76547

A gated, high voltage iontophoresis system with accurate current monitoring.

B M Smith, B J Hoffer.   

Abstract

A circuit description for a multichannel constant current source for microiontophoresis is presented. The circuitry allows a high voltage compliance, accurate monitoring of ionotophoretic current without use of batteries, and automatic current neutralization. Drugs may be applied either manually or with an external digital gate signal. A monitor to detect current unbalance greater than a preselected level is also included.

Mesh:

Year:  1978        PMID: 76547     DOI: 10.1016/0013-4694(78)90316-4

Source DB:  PubMed          Journal:  Electroencephalogr Clin Neurophysiol        ISSN: 0013-4694