Literature DB >> 7651122

A Nyquist modulated echo-to-view mapping scheme for fast spin-echo imaging.

P J Keller1, J E Heiserman, E K Fram, S D Rand, B P Drayer.   

Abstract

A new scheme for assignment of echoes to views in fast spin-echo imaging was developed. The scheme places early and late echoes in alternating lines of the periphery of k-space; continued alternation of echoes arising closer to the middle of the echo train encodes the central portion of k-space. The scheme has two effects: a) The echo-to-echo signal decay that usually gives rise to multiple faint ghosts, now yields a single Nyquist ghost that is removed by phase over-sampling. b) Mapping earlier echoes (and thus more energy) to the periphery of k-space yields sharper edges.

Mesh:

Year:  1995        PMID: 7651122     DOI: 10.1002/mrm.1910330615

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  1 in total

1.  High-resolution fast spin echo imaging of the human brain at 4.7 T: implementation and sequence characteristics.

Authors:  David L Thomas; Enrico De Vita; Steven Roberts; Robert Turner; Tarek A Yousry; Roger J Ordidge