| Literature DB >> 7634875 |
J L Hankinson1, M S Filios, K B Kinsley, E L Petsonk.
Abstract
The accuracy and instrument variability of the MiniWright (Clement Clarke) peak expiratory flow (PEF) meter was determined with 6 of the 24 American Thoracic Society's (ATS) standard waveforms using a mechanical pump. Both room air and air heated to 37 degrees C and saturated with water vapor were used. In addition, MiniWright-determined PEF measurements were compared with those obtained using a dry rolling-seal spirometer (Ohio No. 822; Ohio Medical Products; Madison, Wis) from 75 subjects on 2 different days. The MiniWright average coefficient of variation within a waveform was found to be 2.8%. Results using heated and humidified air (body temperature, ambient pressure, and saturated with water: body conditions) were 2.5% lower than those obtained using room air. Comparisons with mechanically simulated PEF and with spirometry-determined peak flow in 75 human subjects showed that MiniWright meters over-estimated flows at lower flow rates and slightly under-estimated flows at higher flow rates. These results suggest that the new "mechanical PEF" MiniWright scale should be used instead of the "traditional" MiniWright scale.Entities:
Mesh:
Year: 1995 PMID: 7634875 DOI: 10.1378/chest.108.2.407
Source DB: PubMed Journal: Chest ISSN: 0012-3692 Impact factor: 9.410