Literature DB >> 7634606

Hydroxyapatite-coated hip prostheses. Early results from an international study.

A J Tonino1, L Romanini, P Rossi, M Borroni, F Greco, C Garcia-Araujo, L Garcia-Dihinx, A Murcia-Mazòn, W Hein, J Anderson.   

Abstract

A hydroxyapatite-coated hip prosthesis designed to stimulate proximal femoral stress transfer was studied in 222 patients undergoing primary total hip arthroplasty with a minimum followup of 2 years. The femoral component was a roughened titanium alloy with a 50-mu coating of hydroxyapatite applied to the proximal third. A proximal press-fit technique was applied with distal femoral over-reaming as standard procedure. The hydroxyapatite-coated acetabulum was of hemispherical design. Analysis of the clinical results showed a mean Merle D'Aubigné score of 16.7 points at 6 months and 17.4 points at 2 years. Mild residual thigh pain of unknown origin was reported in 3.6% of patients at 2 years. Radiographic evaluation showed complete acetabular and proximal femoral osseointegration without any loosening. No radiolucency or reactive line formation was observed around the hydroxyapatite-coated areas of the prostheses. Femoral bone densification occurred at the coated-uncoated transition zone (54%); 4% to 6% of cases showed cortical hypertrophy at the uncoated area of the stem. Reactive lines were seen around the distal stem in 52% of patients after 2 years. The clinical outcome was excellent, and the radiographic data pointed mainly to proximal femoral stress transfer.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7634606

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  6 in total

1.  Bilayered calcium phosphate coating to promote osseointegration of a femoral stem prosthesis.

Authors:  E Goyenvalle; N J M Guyen; E Aguado; N Passuti; G Daculsi
Journal:  J Mater Sci Mater Med       Date:  2003-03       Impact factor: 3.896

2.  THA with the ABG I prosthesis at 15 years. Excellent survival with minimal osteolysis.

Authors:  P N Baker; I A McMurtry; G Chuter; A Port; J Anderson
Journal:  Clin Orthop Relat Res       Date:  2010-07       Impact factor: 4.176

3.  A stature-specific concept for uncemented, primary total hip arthroplasty.

Authors:  Georg W Omlor; Hannah Ullrich; Knut Krahmer; Alexander Jung; Günther Aldinger; Peter Aldinger
Journal:  Acta Orthop       Date:  2010-02       Impact factor: 3.717

4.  Periprosthetic bone loss after insertion of an uncemented, customized femoral stem and an uncemented anatomical stem. A randomized DXA study with 5-year follow-up.

Authors:  Mona Nysted; Pål Benum; Jomar Klaksvik; Olav Foss; Arild Aamodt
Journal:  Acta Orthop       Date:  2011-06-14       Impact factor: 3.717

5.  The Effect of Hydroxyapatite Coating on Long-term Results of Total Hip Arthroplasty with Hydroxyapatite-coated Anatomic Femoral Stem.

Authors:  Young-Yool Chung; Chae-Hyun Im; Dae-Hee Kim; Ju-Yeong Heo; Young-Jae Jang
Journal:  Hip Pelvis       Date:  2014-09-29

6.  Small and similar amounts of micromotion in an anatomical stem and a customized cementless femoral stem in regular-shaped femurs. A 5-year follow-up randomized RSA study.

Authors:  Mona Nysted; Olav A Foss; Jomar Klaksvik; Pål Benum; Kristin Haugan; Otto Schnell Husby; Arild Aamodt
Journal:  Acta Orthop       Date:  2014-04       Impact factor: 3.717

  6 in total