Comparison of two anti-hepatitis C virus enzyme-linked immunosorbent assays.

H Vrielink1, H L Zaaijer, H W Reesink, P N Lelie, C L van der Poel.   

Abstract

BACKGROUND: Third-generation anti-hepatitis C virus (HCV) enzyme-linked immunosorbent assays (ELISA) are now implemented in most laboratories in Europe, but have not yet been fully implemented in the United States. STUDY DESIGN AND METHODS: Two ELISAs (Ortho 3.0 and Ortho 2.0, Ortho Diagnostics, Raritan, NJ) were compared by tests on various serum panels: A) blood donor samples (n = 530) that tested positive in first- or second-generation anti-HCV ELISA; B) samples from persons with chronic non-A, non-B hepatitis (n = 185); C) samples from multiply transfused patients (n = 79); D) samples from patients on hemodialysis (n = 473); and E) samples from Dutch random blood donors (n = 2153).
RESULTS: In panels A, B, and C, 247 (100%) of 247 polymerase chain reaction (PCR)-positive and 278 (100%) of 278 second-generation recombinant immunoblot assay (RIBA-2)-positive specimens were detected by Ortho 2.0 and 3.0 (sensitivity, 100%). In the sera of panel D, used to represent a group of patients with a high risk for HCV, no additional positives were found by Ortho 3.0. In panel E, of 2153 blood donor samples, 2 (0.1%) were positive in Ortho 2.0 and 8 (0.4%) in Ortho 3.0. Two samples that were positive in both Ortho 2.0 and 3.0 were also positive in RIBA-2; one was positive on PCR. From the 6 remaining Ortho 3.0-positive (Ortho 2.0-negative) samples, 1 was positive in RIBA-2 (isolated anti-c100) and 3 were positive in third-generation RIBA (1/3 isolated anti-c100, 2/3 isolated NS5). All 6 samples were PCR negative. In first-time donors, no difference in specificity was found.
CONCLUSION: The sensitivity and specificity of the Ortho 3.0 ELISA are comparable to those of the Ortho 2.0 ELISA.

Entities:  

Mesh:

Year:  1995        PMID: 7631395     DOI: 10.1046/j.1537-2995.1995.35795357885.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  2 in total

1.  PHS guideline for reducing human immunodeficiency virus, hepatitis B virus, and hepatitis C virus transmission through organ transplantation.

Authors:  Debbie L Seem; Ingi Lee; Craig A Umscheid; Matthew J Kuehnert
Journal:  Public Health Rep       Date:  2013-07       Impact factor: 2.792

2.  Automated RIBA hepatitis C virus (HCV) strip immunoblot assay for reproducible HCV diagnosis.

Authors:  P Martin; F Fabrizi; V Dixit; S Quan; M Brezina; E Kaufman; K Sra; R DiNello; A Polito; G Gitnick
Journal:  J Clin Microbiol       Date:  1998-02       Impact factor: 5.948

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.