Evaluation of the Technicon Immuno I random access immunoassay analyzer and calculation of pediatric reference ranges for endocrine tests, T-uptake, and ferritin.

J N Murthy1, J M Hicks, S J Soldin.   

Abstract

OBJECTIVES: Evaluation of the precision, accuracy, and user-friendliness of the Technician Immuno I. Calculation of pediatric reference ranges for ferritin and endocrine tests run on Immuno I.
METHODS: Precision and accuracy were measured using controls and method comparison studies. Pediatric reference ranges were calculated by comparing the Immuno I results for 100 patients with those of the Abbott IMx and TDx. The regression equation obtained was then used to convert the IMx and TDx reference ranges to reference ranges for the Immuno I.
RESULTS: The Immuno I provided both accurate and precise measurement of drugs and endocrine hormones. Pediatric reference ranges were obtained for ferritin and all endocrine tests.
CONCLUSION: The Immuno I is user-friendly and provides reliable measurement of both the drugs tested and endocrine tests on a micro-sample. Reagent and curve stability are excellent.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7628078     DOI: 10.1016/0009-9120(94)00080-f

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  5 in total

1.  Association between neonatal iron overload and early human brain development in premature infants.

Authors:  Sanjiv B Amin; Gary Myers; Hongyue Wang
Journal:  Early Hum Dev       Date:  2012-02-18       Impact factor: 2.079

2.  Comparison of FT4 with log TSH on the Abbott Architect ci8200: Pediatric reference intervals for free thyroxine and thyroid-stimulating hormone.

Authors:  Steven J Soldin; Luke L Cheng; Lisa Y Lam; Alice Werner; Alexander D Le; Offie P Soldin
Journal:  Clin Chim Acta       Date:  2009-11-24       Impact factor: 3.786

3.  Pediatric reference ranges for zinc protoporphyrin.

Authors:  Offie Porat Soldin; Maureen Miller; Steven J Soldin
Journal:  Clin Biochem       Date:  2003-02       Impact factor: 3.281

4.  Pediatric reference intervals for free thyroxine and free triiodothyronine.

Authors:  Offie P Soldin; Megan Jang; Tiedong Guo; Steven J Soldin
Journal:  Thyroid       Date:  2009-07       Impact factor: 6.568

Review 5.  Performance and comparability of laboratory methods for measuring ferritin concentrations in human serum or plasma: A systematic review and meta-analysis.

Authors:  Maria N Garcia-Casal; Juan P Peña-Rosas; Eloisa Urrechaga; Jesus F Escanero; Junsheng Huo; Ricardo X Martinez; Lucero Lopez-Perez
Journal:  PLoS One       Date:  2018-05-03       Impact factor: 3.240

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.