On the detection of QRS variations in the ECG.

G R Shaw1, P Savard.   

Abstract

Detection of subtle beat-to-beat variations in the morphology of the ECG is complicated by the effects of alignment errors and respiration. A method of directly estimating the alignment error (trigger jitter) from an ECG is derived by relating the variance to the squared slope of the averaged QRS complex. Results based on recordings obtained from 12 normal subjects and alignment performed by the cross-correlation method showed that the alignment errors were dependent upon the choice of the alignment channel, with the best distribution of the errors occurring when alignment was based on the vector magnitude of the three orthogonal leads. The estimated average alignment errors ranged from 0.33-0.42 ms, which were near the optimal value of 0.29 ms based on the sampling rate of 1000 samples/s. It was shown that the effects of respiration could be reduced by normalizing the amplitude of the QRS complexes. It was also estimated that a significant proportion of the variation (0.54-0.67) in the normalized ECG's could be attributed to alignment errors and noise.

Mesh:

Year:  1995        PMID: 7622158     DOI: 10.1109/10.391174

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  2 in total

1.  Sampling rate and the estimation of ensemble variability for repetitive signals.

Authors:  P Laguna; L Sörnmo
Journal:  Med Biol Eng Comput       Date:  2000-09       Impact factor: 2.602

2.  A novel low-complexity post-processing algorithm for precise QRS localization.

Authors:  Pedro Fonseca; Ronald M Aarts; Jérôme Foussier; Xi Long
Journal:  Springerplus       Date:  2014-07-25
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.