Comparative analysis of power-line interference between two- or three-electrode biopotential amplifiers.

D E Wood1, D J Ewins, W Balachandran.   

Abstract

Differential input amplifiers are commonly used in the measurement of biopotential signals. This is because of their ability to reject common-mode signals, for example power-line interference. Rejection is further enhanced by referencing the subject to the instrumentation system. One such referencing technique uses a third subject electrode connected either to circuit common or to ground. However, these techniques are rarely explained or quantified in the literature. Four recording configurations are examined, isolated and non-isolated amplifiers with two and with three electrodes, and existing models are extended. From a brief assessment of power-line interference sources, it is concluded that only displacement currents induced into the leads and the body need be considered. The effect of these on the four configurations is analysed using SPICE. Results show that interference from displacement currents induced into the leads and into the body is dependent upon the recording configuration used and is significantly lower in three-electrode than in two-electrode amplifiers. Isolation also reduces levels of interference. It is concluded that the choice of referencing may affect interference levels on the biopotential signal significantly. Experimental and further analytical work is planned to confirm this.

Mesh:

Year:  1995        PMID: 7616784     DOI: 10.1007/BF02522948

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  14 in total

1.  The isolation mode rejection ratio in bioelectric amplifiers.

Authors:  A C Metting van Rijn; A Peper; C A Grimbergen
Journal:  IEEE Trans Biomed Eng       Date:  1991-11       Impact factor: 4.538

2.  Miniature skin-mounted preamplifier for measurement of surface electromyographic potentials.

Authors:  S W Johnson; P A Lynn; J S Miller; G A Reed
Journal:  Med Biol Eng Comput       Date:  1977-11       Impact factor: 2.602

3.  Skin impedance from 1 Hz to 1 MHz.

Authors:  J Rosell; J Colominas; P Riu; R Pallas-Areny; J G Webster
Journal:  IEEE Trans Biomed Eng       Date:  1988-08       Impact factor: 4.538

4.  On the reduction of interference due to common mode voltage in two-electrode biopotential amplifiers.

Authors:  R Pallás-Areny
Journal:  IEEE Trans Biomed Eng       Date:  1986-11       Impact factor: 4.538

5.  60-HZ interference in electrocardiography.

Authors:  J C Huhta; J G Webster
Journal:  IEEE Trans Biomed Eng       Date:  1973-03       Impact factor: 4.538

6.  Interference-rejection characteristics of biopotential amplifiers: a comparative analysis.

Authors:  R Pallás-Areny
Journal:  IEEE Trans Biomed Eng       Date:  1988-11       Impact factor: 4.538

7.  Mains interference in ECG recording.

Authors:  I Marshall; J M Neilson
Journal:  J Med Eng Technol       Date:  1984 Jul-Aug

8.  Driven-right-leg circuit design.

Authors:  B B Winter; J G Webster
Journal:  IEEE Trans Biomed Eng       Date:  1983-01       Impact factor: 4.538

9.  Reduction of interference due to common mode voltage in biopotential amplifiers.

Authors:  B B Winter; J G Webster
Journal:  IEEE Trans Biomed Eng       Date:  1983-01       Impact factor: 4.538

10.  How to prevent 60-Hz interference.

Authors:  J G Webster
Journal:  Med Instrum       Date:  1979 Sep-Oct
View more
  5 in total

1.  Two-electrode biopotential amplifier with current-driven inputs.

Authors:  D Dobrev; I Daskalov
Journal:  Med Biol Eng Comput       Date:  2002-01       Impact factor: 2.602

2.  Estimation of stray coupling capacitances in biopotential measurements.

Authors:  Marcelo Haberman; Augusto Cassino; Enrique Spinelli
Journal:  Med Biol Eng Comput       Date:  2011-07-20       Impact factor: 2.602

3.  Biopotential amplifier for simultaneous operation with biomagnetic instruments.

Authors:  J Virtanen; L Parkkonen; R J Ilmoniemi; E Pekkonen; R Näätänen
Journal:  Med Biol Eng Comput       Date:  1997-07       Impact factor: 2.602

4.  Teaching Essential EMG Theory to Kinesiologists and Physical Therapists Using Analogies Visual Descriptions, and Qualitative Analysis of Biophysical Concepts.

Authors:  David A Gabriel
Journal:  Sensors (Basel)       Date:  2022-08-30       Impact factor: 3.847

5.  Two-Electrode ECG for Ambulatory Monitoring with Minimal Hardware Complexity.

Authors:  Branko Babusiak; Stefan Borik; Maros Smondrk
Journal:  Sensors (Basel)       Date:  2020-04-22       Impact factor: 3.576

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.