The effect of haemolysis on blood glucose meter measurement.

E S Kilpatrick1, A G Rumley, C N Rumley.   

Abstract

A study was performed to assess the effect of varying degrees of sample haemolysis on the measurement of blood glucose by the Accutrend, Companion 2, ExacTech, Glucometer II, Glucometer 4, One Touch II, and Reflolux II blood glucose meters. Fresh venous blood was sonicated to induce complete haemolysis and then added in increasing proportions to homologous untreated blood to obtain nine samples with free haemoglobin concentrations up to 50 g l-1. The Accutrend meter showed the only significant (p < 0.05) linear relationship to degree of haemolysis (r = 0.988, p < 0.0001). For every 7% of red cells lysed, the Accutrend value increased by 15%. All other meters gave results which were within 15% of the non-haemolysed value. However, extreme (100%) haemolysis not only affected the Accutrend (glucose value 108% greater than reference) but also the ExacTech (+98%), the Glucometer II (-32%), and the Companion 2 (-41%). Thus, unwitting use of a haemolysed sample to measure whole blood glucose may, with the Accutrend in particular, lead to erroneous results.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7600751     DOI: 10.1111/j.1464-5491.1995.tb00489.x

Source DB:  PubMed          Journal:  Diabet Med        ISSN: 0742-3071            Impact factor:   4.359


  3 in total

1.  Blood collection for biochemical analysis in adult zebrafish.

Authors:  Gabriela L Pedroso; Thais O Hammes; Thayssa D C Escobar; Laisa B Fracasso; Luiz Felipe Forgiarini; Themis R da Silveira
Journal:  J Vis Exp       Date:  2012-05-26       Impact factor: 1.355

2.  The hitchhiker's guide to research in clinical biochemistry.

Authors:  Eric S Kilpatrick
Journal:  Clin Biochem Rev       Date:  2010-02

3.  Blood sugar measurement in zebrafish reveals dynamics of glucose homeostasis.

Authors:  Stefani C Eames; Louis H Philipson; Victoria E Prince; Mary D Kinkel
Journal:  Zebrafish       Date:  2010-06       Impact factor: 1.985

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.