Literature DB >> 7593103

Fixation strength of a biodegradable screw in anterior cruciate ligament reconstruction.

P Kousa1, T L Järvinen, T Pohjonen, P Kannus, M Kotikoski, M Järvinen.   

Abstract

We compared the strength of fixation of a biodegradable screw with that of two metal screws in a bone-patellar-tendon-bone (BPTB) graft in the bovine knee. We used 33 fresh BPTB specimens with a circular tibial bone plug of 9 mm in diameter which were anchored in a tibial metaphyseal bone tunnel with either an interference screw (n = 11), an AO cancellous screw (n = 11) or a fibrillated, self-reinforced biodegradable poly-L-lactide screw (n = 11). The mean force to failure (+/- SD) in the three groups was 1358 +/- 348 N, 1081 +/- 331 N and 1211 +/- 362 N, respectively. There was no significant difference in the groups with regard to the linear load or the elastic moduli of the fixation. We conclude that the biodegradable screw is as good as either of the two metal screws in the fixation of a BPTB graft in the bovine knee and can be recommended for ACL reconstruction using this type of graft.

Entities:  

Mesh:

Year:  1995        PMID: 7593103

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  10 in total

1.  Analysis of initial fixation strength of press-fit fixation technique in anterior cruciate ligament reconstruction. A comparative study with titanium and bioabsorbable interference screw using porcine lower limb.

Authors:  Myung Chul Lee; Hyunchul Jo; Tae-Soo Bae; Jin Dae Jang; Sang Cheol Seong
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-03-06       Impact factor: 4.342

2.  Equal tension in all four strands of a semitendinosus graft with a low profile tibial fixation device (Cobra LFD).

Authors:  S Winge; P Dunn; B K O Engström
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2005-10-06       Impact factor: 4.342

3.  Poly-L-lactic acid - hydroxyapatite (PLLA-HA) bioabsorbable interference screws for tibial graft fixation in anterior cruciate ligament (ACL) reconstruction surgery: MR evaluation of osteointegration and degradation features.

Authors:  L Macarini; P Milillo; A Mocci; R Vinci; G C Ettorre
Journal:  Radiol Med       Date:  2008-10-25       Impact factor: 3.469

Review 4.  Graft choice and graft fixation in PCL reconstruction.

Authors:  Jürgen Höher; Sven Scheffler; Andreas Weiler
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2003-08-26       Impact factor: 4.342

5.  Tibial tunnel widening after anterior cruciate ligament reconstructions with hamstring tendons using Rigidfix femoral fixation and Intrafix tibial fixation.

Authors:  Nam-Hong Choi; Jung-Hoon Lee; Kyung-Mo Son; Brian N Victoroff
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-01       Impact factor: 4.342

6.  Biomechanical characterization of double-bundle femoral press-fit fixation techniques.

Authors:  M Ettinger; C Haasper; S Hankemeier; C Hurschler; D Breitmeier; C Krettek; M Jagodzinski
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-07-14       Impact factor: 4.342

7.  Fixation strength of biocomposite wedge interference screw in ACL reconstruction: effect of screw length and tunnel/screw ratio. A controlled laboratory study.

Authors:  Antonio Herrera; Fernando Martínez; Daniel Iglesias; José Cegoñino; Elena Ibarz; Luis Gracia
Journal:  BMC Musculoskelet Disord       Date:  2010-06-30       Impact factor: 2.362

8.  Resorbable screw and sheath versus resorbable interference screw and staples for ACL reconstruction: a comparison of two tibial fixation methods.

Authors:  Christian Carulli; Fabrizio Matassi; Stefano Soderi; Luigi Sirleo; Giovanni Munz; Massimo Innocenti
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-04-27       Impact factor: 4.342

9.  Biomechanical evaluation of a new cross-pin technique for the fixation of different sized bone-patellar tendon-bone grafts.

Authors:  Thore Zantop; Barbara Welbers; Andre Weimann; Markus Rümmler; Jürgen Hedderich; Volker Musahl; Wolf Petersen
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2004-06-19       Impact factor: 4.342

10.  Fixation strength of a novel bioabsorbable expansion bolt for patellar tendon bone graft fixation: an experimental study in calf tibial bone.

Authors:  Stefan Piltz; Patrick Strunk; Ludger Meyer; Wolfgang Plitz; Guenter Lob
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2004-03-24       Impact factor: 4.342

  10 in total