The aim of the present study was to evaluate the single needle (SN) procedure for plateletpheresis with the AS 104 blood cell separator. One hundred fifty-seven platelet collections with different protocols were carried out and analysed with respect to platelet yield, separation efficiency, leukocyte contamination, anticoagulant infused, and donation time. SN plateletpheresis resulted in a mean platelet yield of 2.6-3.1 x 10(11) with separation efficiencies of 49.2-57.4%. The mean leukocyte contamination ranged from 2.8 to 40.5 x 10(6). Parameters of product quality were investigated in 14 platelet concentrates (PC). Determination of glucose, lactate, lactate dehydrogenase (LDH), and morphology score showed good quality of the PC.Entities:
Mesh:
Year: 1995 PMID: 7592524 DOI: 10.1002/jca.2920100208
Source DB: PubMed Journal: J Clin Apher ISSN: 0733-2459 Impact factor: 2.821