Transferability of blood lead determinations by furnace atomic absorption spectrophotometry and continuum background correction.

H Qiao1, P J Parsons, W Slavin.   

Abstract

We have examined and proved feasible the transfer of a method for blood lead determination, developed and optimized for a Zeeman-corrected instrument, to a continuum-corrected furnace atomic absorption spectrophotometer. Numerous reference materials analyzed with the continuum-corrected instrument gave results within 10 micrograms/L (0.05 mumol/L) at low values and varied by < 6% at values > 200 micrograms/L (0.97 mumol/L). Forty-four routine human blood specimens were analyzed by the same method with both continuum- and Zeeman-corrected instrumentation, and gave results that agreed within about the same limits as found with reference materials. The day-to-day precision was about 1/5 the accuracy results. The detection limit was approximately 5 micrograms/L (0.025 mumol/L).

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7586516

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  2 in total

1.  Factors influencing the difference between maternal and cord blood lead.

Authors:  E W Harville; I Hertz-Picciotto; M Schramm; M Watt-Morse; K Chantala; J Osterloh; P J Parsons; W Rogan
Journal:  Occup Environ Med       Date:  2005-04       Impact factor: 4.402

2.  Dose-Response Relationship between Cumulative Occupational Lead Exposure and the Associated Health Damages: A 20-Year Cohort Study of a Smelter in China.

Authors:  Yue Wu; Jun-Ming Gu; Yun Huang; Yan-Ying Duan; Rui-Xue Huang; Jian-An Hu
Journal:  Int J Environ Res Public Health       Date:  2016-03-16       Impact factor: 3.390

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.