Fluctuations in syringe-pump infusions: association with blood pressure variations in infants.

D F Capes1, K R Dunster, V B Sunderland, D McMillan, P B Colditz, C McDonald.   

Abstract

Flow continuity of two brands of syringe pumps and four brands of syringes was studied as a possible cause of hemodynamic fluctuations observed in neonates. Cyclical fluctuations were observed in the blood pressure of 14 neonates receiving dopamine infusions by syringe pump at flow rates from 0.2 to 1 mL/hr. Atom 235 and IVAC 770 pumps and various sizes of Terumo, Becton Dickinson, Omnifix, and IVAC syringes were evaluated. Flow continuity was assessed by using a gravimetric technique. The force needed to initiate and maintain syringe plunger motion was also measured. Noncontinuous flow was encountered most commonly with Terumo syringes, which delivered boluses at regular intervals at flow rates up to 5 mL/hr. The interval was dependent on flow rate and was similar to the time between the blood pressure fluctuations observed clinically. The syringe plunger force exhibited regular fluctuations indicative of the plunger sticking, and simultaneous measurement of flow established a direct temporal relationship with boluses. The other syringes tested did not exhibit such fluctuations. No differences were found between the two syringe pumps. Syringe plunger sticking, resulting in intermittent boluses and potential blood pressure fluctuations, may occur at low flow rates and with certain syringe brands. This appeared to be the cause of hemodynamic fluctuations in neonates receiving dopamine infusions.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7583827     DOI: 10.1093/ajhp/52.15.1646

Source DB:  PubMed          Journal:  Am J Health Syst Pharm        ISSN: 1079-2082            Impact factor:   2.637


  2 in total

1.  Delay and stability of central venous administration of norepinephrine in children: a bench study.

Authors:  Mehdi Oualha; Mikaël Capelo; Odile Spreux-Varoquaux; Isabelle Drouet-Chaillou; Jean-Marc Tréluyer; Philippe Hubert; Fabrice Lesage
Journal:  Eur J Pediatr       Date:  2013-08-11       Impact factor: 3.183

2.  Extreme hemodynamic fluctuations: Importance of understanding the principles of syringe pump function.

Authors:  Chakrabarti Dhritiman; Byrappa Vinay
Journal:  Indian J Crit Care Med       Date:  2015-04