Evaluation of the Paratrend 7 intravascular blood gas monitor during cardiac surgery: comparison with the C4000 in-line blood gas monitor during cardiopulmonary bypass.

B Venkatesh1, T H Clutton-Brock, S P Hendry.   

Abstract

OBJECTIVE: To evaluate the performance of the Paratrend 7 intravascular blood gas monitor (Biomedical Sensors, High Wycombe, UK, Ltd) during cardiac surgery and compare it with that of an in-line blood gas monitor placed in the arterial limb of an extracorporeal circuit during cardiopulmonary bypass.
DESIGN: A prospective study. Consecutive patient enrolment.
SETTING: In the cardiac surgical intensive care units at a tertiary referral center. INTERVENTION: Insertion of the Paratrend 7 intravascular sensor through the radial arterial catheter after induction of anesthesia.
MEASUREMENTS AND MAIN RESULTS: Simultaneous measurements of pH, PCO2, and PO2 were made from the sensor and the blood gas analyzer, and the bias and precision were calculated on all the measured parameters. The bias and precision of the intravascular sensor during bypass for pH, PCO2, and PO2 were 0.01 and 0.06 pH units, 0.5 and 2.5 mmHg (2% and 8%), and 3 and 45 torr (0.5% and 14%), respectively. The bias and precision for the prebypass and the postbypass phases were comparable. The bias and precision of the extracorporeal monitor for pH, PCO2, and PO2 were 0.04 and 0.1 pH units, -0.3 and 4 mmHg (-1% and 15%) and 8 and 48 mmHg (4 and 18%), respectively. There were no instances of any complications attributable to the intravascular sensor.
CONCLUSIONS: The intravascular sensor used in this study functioned well during cardiopulmonary bypass and the postbypass phase. The performance of the intravascular sensor was better than the in-line blood gas monitor during cardiopulmonary bypass.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7579111     DOI: 10.1016/s1053-0770(05)80096-5

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  4 in total

Review 1.  Continuous intra-arterial blood gas monitoring.

Authors:  B Venkatesh; S P Hendry
Journal:  Intensive Care Med       Date:  1996-08       Impact factor: 17.440

2.  Optical Sensing and Imaging of pH Values: Spectroscopies, Materials, and Applications.

Authors:  Andreas Steinegger; Otto S Wolfbeis; Sergey M Borisov
Journal:  Chem Rev       Date:  2020-11-04       Impact factor: 60.622

3.  Off-pump coronary artery bypass grafting in a patient with paroxysmal nocturnal hemoglobinuria.

Authors:  Yutaka Makino; Yukio Suto; Tatsuya Murakami
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2004-10

4.  Assessment of a continuous blood gas monitoring system in animals during circulatory stress.

Authors:  Sandro Gelsomino; Roberto Lorusso; Ugolino Livi; Stefano Romagnoli; Salvatore Mario Romano; Rocco Carella; Fabiana Lucà; Giuseppe Billè; Francesco Matteucci; Attilio Renzulli; Gil Bolotin; Giuseppe De Cicco; Pierluigi Stefàno; Jos Maessen; Gian Franco Gensini
Journal:  BMC Anesthesiol       Date:  2011-01-11       Impact factor: 2.217

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.