M Sandborg1, J O Christoffersson, G A Carlsson, T Almén, D R Dance.
Abstract
A Monte Carlo computational model of the imaging chain has been used to investigate the performance of x-ray contrast agents with atomic number, Z, 53 < or = Z < or = 90 with respect to physical image quality descriptors (contrast and signal to noise ratio, SNR) and patient mean absorbed dose. Contrast agents of equal molar concentrations were used within a water slab (simulating the patient). The imaging conditions were chosen to represent adult and paediatric examinations. For all tube potentials studied (40-140 kV), the contrast agents with the highest atomic numbers (bismuth and thorium) gave the highest contrast. In analogue screen-film imaging, several other contrast agents could produce a higher image contrast than iodine in a limited range of tube potentials. This advantage could alternatively be effected as a reduced amount of administered contrast agent, or as a reduced mean absorbed dose in the patient. In digital imaging, a lower mean absorbed dose for a constant SNR than that with iodine can be achieved for ranges of tube potentials and contrast agents. Bismuth and thorium yield a lower dose than iodine at all studied tube potentials. Gadolinium and erbium could alternatively be used at a broad range of tube potentials above 90 kV with a dose penalty of only 5-20%.Entities:
Mesh:
Substances:
Year: 1995 PMID: 7568378 DOI: 10.1088/0031-9155/40/7/005
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609