Literature DB >> 7567667

Persistent hemolysis after transcatheter occlusion of a patent ductus arteriosus: surgical ligation of the duct over the occlusion device.

J C Chisholm1, A P Salmon, B R Keeton, S A Webber, J L Monro.   

Abstract

A 14-month-old child underwent routine transcatheter closure of her patent ductus arteriosus (PDA) using the Rashkind double-umbrella device. She developed severe persistent hemolysis in association with residual ductal flow. The problem was managed by the previously unreported approach of standard surgical ligation of the duct via a left thoracotomy, leaving the occlusion device in situ.

Entities:  

Mesh:

Year:  1995        PMID: 7567667     DOI: 10.1007/BF00794194

Source DB:  PubMed          Journal:  Pediatr Cardiol        ISSN: 0172-0643            Impact factor:   1.655


  7 in total

1.  Transcatheter occlusion of persistent arterial duct. Report of The European Registry.

Authors: 
Journal:  Lancet       Date:  1992-10-31       Impact factor: 79.321

2.  Late transcatheter removal of a Rashkind PDA occlusion device for persistent hemolysis using a modified transseptal sheath.

Authors:  R G Grifka; M P O'Laughlin; C E Mullins
Journal:  Cathet Cardiovasc Diagn       Date:  1992-02

3.  Severe haemolysis after percutaneous closure of a ductus arteriosus (arterial duct).

Authors:  E J Ladusans; I Murdoch; J Franciosi
Journal:  Br Heart J       Date:  1989-06

4.  Two-dimensional echocardiographic and color flow Doppler evaluation of ductal occlusion with the Rashkind prosthesis.

Authors:  N N Musewe; L N Benson; J F Smallhorn; R M Freedom
Journal:  Circulation       Date:  1989-12       Impact factor: 29.690

5.  Transcatheter occlusion of the persistently patent ductus arteriosus. Forty-month follow-up and prevalence of residual shunting.

Authors:  M C Hosking; L N Benson; N Musewe; J D Dyck; R M Freedom
Journal:  Circulation       Date:  1991-12       Impact factor: 29.690

6.  Nonsurgical closure of patent ductus arteriosus: clinical application of the Rashkind PDA Occluder System.

Authors:  W J Rashkind; C E Mullins; W E Hellenbrand; M A Tait
Journal:  Circulation       Date:  1987-03       Impact factor: 29.690

7.  Experience with 205 procedures of transcatheter closure of ductus arteriosus in 182 patients, with special reference to residual shunts and long-term follow-up.

Authors:  M A Ali Khan; S al Yousef; C E Mullins; W Sawyer
Journal:  J Thorac Cardiovasc Surg       Date:  1992-12       Impact factor: 5.209