Literature DB >> 7567203

Functional residual capacity (FRC) measurements by plethysmography and helium dilution in normal infants.

K S McCoy1, R G Castile, E D Allen, D A Filbrun, R L Flucke, E Bar-Yishay.   

Abstract

Comparative measurements of functional residual capacity (FRC) made by plethysmography (FRCpleth) and by helium dilution (FRCHe) were obtained on 27 infants and young children without known pulmonary disease (14 males, 13 females; 4 weeks-26 months; mean age 32.2 weeks) while under chloral hydrate sedation. Clinical histories, clinical examinations, and pulmonary functions were normal for all members of the group. FRCpleth, whether measured near end expiration (EE) or near end inspiration (EI), and corrected to mean expiratory levels of at least 3 breathing cycles, was consistently and significantly greater than FRCHe. Comparative values for mean (+/- standard deviation) were FRCpleth EE, 182.0 (+/- 79.7) mL and FRCpleth El, 171.8 (+/- 77.4) mL vs. FRCHe, 154 (+/- 72.2) mL, P < 0.0001 and P < 0.005, respectively. Normalizing values by weight, FRCpleth EE was 23.8 mL/kg (+/- 5.3) vs. FRCHe, 20.2 (+/- 4.7) mL/kg, mean (+/- standard deviation). The difference between FRCpleth and FRCHe, expressed as FRCpleth - FRCHe/FRCpleth x 100, was 9% for occlusions at end inspiration and 16% for occlusions at end expiration. The following equations describe our FRC results in relation to length: In (FRCHe) = 2.74 x ln (length) - 6.53 r2 = 0.781 slope = 2.74 +/- 0.29 SE Y intercept = 6.53 +/- 1.12 SE ln (FRCPleth EI) = 2.69 x ln (length) - 6.21 r2 = 0.752 slope = 2.69 +/- 0.31 SE Y intercept = 6.21 +/- 1.29 SE The difference between FRCpleth and FRCHe was more marked when occlusions were performed at end expiration than at end inspiration.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7567203     DOI: 10.1002/ppul.1950190507

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  3 in total

Review 1.  Endpoints for clinical trials in young children with cystic fibrosis.

Authors:  Stephanie D Davis; Alan S Brody; Mary J Emond; Lyndia C Brumback; Margaret Rosenfeld
Journal:  Proc Am Thorac Soc       Date:  2007-08-01

2.  Measurement of lung volumes.

Authors:  Theodore G Liou; Richard E Kanner
Journal:  Clin Rev Allergy Immunol       Date:  2009-04-02       Impact factor: 8.667

3.  Hyperinflation is associated with increased respiratory rate and is a more sensitive measure of cystic fibrosis lung disease during infancy compared to forced expiratory measures.

Authors:  Heather N Muston; James E Slaven; Christina Tiller; Charles Clem; Thomas W Ferkol; Sarath Ranganathan; Stephanie D Davis; Clement L Ren
Journal:  Pediatr Pulmonol       Date:  2021-06-23
  3 in total