Nitric oxide administration using constant-flow ventilation.

J W Skimming1, S Cassin, P B Blanch.   

Abstract

Nitric oxide (NO) gas is known as both a vasodilator and a toxin. It can react with oxygen to form compounds more toxic than itself, such as nitrogen dioxide (NO2). The reactions are time dependent; thus, infusing NO into breathing circuits as close to ventilated subjects as possible may help minimize toxic byproduct exposure. Unfortunately, flow rates commonly used with mechanical ventilation favor laminar gas flow (streaming) within the breathing circuits. Streaming could delay mixing of NO with other inhaled gases. This mixing delay may interfere with accurate monitoring and/or delivery of NO. We tested the hypothesis that streaming of NO infused by constant flow into the inspiratory limb of a constant-flow mechanical ventilation system can lead to NO concentration delivery estimate errors. We then compared the NO2 concentrations at the ventilator Y-piece with three different NO mixing methods: blending the gases before they reach the breathing circuit inspiratory limb, infusing NO directly into the breathing circuit inspiratory limb far enough from the Y-piece to ensure thorough mixing, and infusing NO directly into the breathing circuit inspiratory limb immediately before the gases reach an in-line mixing device placed close to the Y-piece. Our results indicate that streaming can lead to NO concentration delivery estimate errors and that these errors can be characterized by measuring NO concentration variations across the inspiratory tubing's luminal diameter. NO2 concentration measured at the ventilator Y-piece were dependent on NO concentrations (p < 0.0001), NO delivery methods (p < 0.0001), and interactions between NO concentrations and NO delivery methods (p < 0.0001). We conclude that gas streaming and toxic byproduct exposure should be considered together when choosing an NO delivery method.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7555121     DOI: 10.1378/chest.108.4.1065

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  2 in total

1.  A European survey of the use of inhaled nitric oxide in the ICU. Working Group on Inhaled NO in the ICU of the European Society of Intensive Care Medicine.

Authors:  S Beloucif; D Payen
Journal:  Intensive Care Med       Date:  1998-08       Impact factor: 17.440

2.  Propagation of nitric oxide pools during controlled mechanical ventilation.

Authors:  J W Skimming; P J Stephan; P B Blanch; M J Banner
Journal:  J Clin Monit Comput       Date:  1998-04       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.