Accuracy of MR phase contrast velocity measurements for unsteady flow.

R Frayne1, D A Steinman, C R Ethier, B K Rutt.   

Abstract

The accuracy of MR phase contrast (PC) velocity measurements for unsteady flow has been quantitatively assessed. Spatially resolved velocity fields were measured in a long straight tube using a gated PC technique, and the resulting MR PC velocity data were compared with velocities derived from the analytic Womersley solution to the Navier-Stokes equations governing fluid flow. The overall root-mean-square (rms) difference between the measured and analytic velocities was 1.6 cm s-1 for nominally sinusoidal flow waveforms with peak velocities ranging from 51.6 cm s-1 to 59.8 cm s-1. This rms difference corresponded to 7.5% of the mean fluid velocity, which is similar to the cited accuracy of approximately 5% for MR PC velocimetry for steady flows. Linear regression between the PC velocity measurements and the velocities obtained using the analytic expression was highly significant (r2 = 0.997) and yielded a slope of 0.998, close to the expected value of 1. We conclude that the gated MR PC velocity measurements in unsteady flow are accurate.

Entities:  

Mesh:

Year:  1995        PMID: 7549205     DOI: 10.1002/jmri.1880050410

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  18 in total

1.  MR-velocity mapping in vascular stents to assess peak systolic velocity. In vitro comparison of various stent designs made of Stainless Steel and Nitinol.

Authors:  Jacqueline van Holten; Patrik Kunz; Paul G H Mulder; Peter M T Pattynama; Hildo J Lamb; Lukas C van Dijk
Journal:  MAGMA       Date:  2002-11       Impact factor: 2.310

2.  Integrated physiological flow simulator and pulse sequence monitoring system for MRI.

Authors:  Pauline Wong; Martin J Graves; David J Lomas
Journal:  Med Biol Eng Comput       Date:  2008-02-22       Impact factor: 2.602

Review 3.  [Diagnosing stroke aetiologies. Morphologic and functional analysis of the aorta and carotid arteries by MRI].

Authors:  A Harloff; M Markl; A Frydrychowicz; J Hennig; C Weiller
Journal:  Nervenarzt       Date:  2009-08       Impact factor: 1.214

4.  Changes in systemic and pulmonary blood flow distribution in normal adult volunteers in response to posture and exercise: a phase contrast magnetic resonance imaging study.

Authors:  Derek T H Wong; Kyong-Jin Lee; Shi-Joon Yoo; George Tomlinson; Lars Grosse-Wortmann
Journal:  J Physiol Sci       Date:  2014-01-03       Impact factor: 2.781

5.  Aqueductal Stroke Volume: Comparisons with Intracranial Pressure Scores in Idiopathic Normal Pressure Hydrocephalus.

Authors:  G Ringstad; K E Emblem; O Geier; N Alperin; P K Eide
Journal:  AJNR Am J Neuroradiol       Date:  2015-05-14       Impact factor: 3.825

6.  Quantitative Flow Imaging in Human Umbilical Vessels In Utero Using Nongated 2D Phase Contrast MRI.

Authors:  Uday Krishnamurthy; Brijesh K Yadav; Pavan K Jella; Ewart Mark Haacke; Edgar Hernandez-Andrade; Swati Mody; Lami Yeo; Sonia S Hassan; Roberto Romero; Jaladhar Neelavalli
Journal:  J Magn Reson Imaging       Date:  2017-12-23       Impact factor: 4.813

7.  Comparison of blood velocity measurements between ultrasound Doppler and accelerated phase-contrast MR angiography in small arteries with disturbed flow.

Authors:  Jingfeng Jiang; Charles Strother; Kevin Johnson; Sara Baker; Dan Consigny; Oliver Wieben; James Zagzebski
Journal:  Phys Med Biol       Date:  2011-02-23       Impact factor: 3.609

8.  Improved Semiautomated 4D Flow MRI Analysis in the Aorta in Patients With Congenital Aortic Valve Anomalies Versus Tricuspid Aortic Valves.

Authors:  Susanne Schnell; Pegah Entezari; Riti J Mahadewia; S Chris Malaisrie; Patrick M McCarthy; Jeremy D Collins; James Carr; Michael Markl
Journal:  J Comput Assist Tomogr       Date:  2016 Jan-Feb       Impact factor: 1.826

9.  Quantification of hemodynamic wall shear stress in patients with bicuspid aortic valve using phase-contrast MRI.

Authors:  Alex J Barker; Craig Lanning; Robin Shandas
Journal:  Ann Biomed Eng       Date:  2009-12-02       Impact factor: 3.934

10.  CSF flow quantification of the cerebral aqueduct in normal volunteers using phase contrast cine MR imaging.

Authors:  Jeong Hyun Lee; Ho Kyu Lee; Jae Kyun Kim; Hyun Jeong Kim; Ji Kang Park; Choong Gon Choi
Journal:  Korean J Radiol       Date:  2004 Apr-Jun       Impact factor: 3.500

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.