BACKGROUND: A simple new device is proposed for safe and very fast epithelial removal of the cornea. This is a rotating plastic brush that removes the corneal epithelium within a few seconds under irrigation, without causing any mechanical damage to the stromal surface.
METHODS: Comparative SEM and TEM studies on rabbit corneas were carried out following epithelial removal by rotating brush and by a Beaver knife blade. Epithelial removal time and reepithelialization time after photorefractive keratectomy were evaluated in a series of 40 human sighted eyes treated with the brush.
RESULTS: The rotating brush-abraded surface was smoother compared to the blade-abraded one. Additionally, the brush provoked no damage to the basal lamina of the rabbit corneal epithelium. In human photorefractive keratectomy, the mean time needed for removal of the corneal epithelium by the rotating brush was only 3 sec (range, 2-5 sec). Reepithelialization time following photorefractive keratectomy did not exceed 3 days.
CONCLUSION: Experimental and preliminary clinical studies suggest that the new rotating plastic brush is a safe and fast method for removing the corneal epithelium.

Entities:  

Mesh:

Year:  1994        PMID: 7528616

Source DB:  PubMed          Journal:  J Refract Corneal Surg        ISSN: 1081-0803


  5 in total

1.  The epithelial flap for photorefractive keratectomy.

Authors:  S Shah; A R Sebai Sarhan; S J Doyle; C T Pillai; H S Dua
Journal:  Br J Ophthalmol       Date:  2001-04       Impact factor: 4.638

2.  Alcohol-assisted versus Mechanical Epithelium Removal in Photorefractive Keratectomy.

Authors:  Mohammad Ghoreishi; Hossein Attarzadeh; Mehdi Tavakoli; Heidar Ali Moini; Alireza Zandi; Amin Masjedi; Akram Rismanchian
Journal:  J Ophthalmic Vis Res       Date:  2010-10

3.  Long term followup of photorefractive keratectomy with adjuvant use of mitomycin C.

Authors:  Vasilios F Diakonis; Vardhaman P Kankariya; George D Kymionis; Georgios Kounis; George Kontadakis; Eirineos Gkenos; Michael A Grentzelos; George Hajithanasis; Sonia H Yoo; Ioannis G Pallikaris
Journal:  J Ophthalmol       Date:  2014-04-29       Impact factor: 1.909

4.  Comparison of mechanical debridement and trans-epithelial myopic photorefractive keratectomy: A contralateral eye study.

Authors:  Siamak Zarei-Ghanavati; Javad Heravian Shandiz; Mojtaba Abrishami; Maliheh Karimpour
Journal:  J Curr Ophthalmol       Date:  2019-01-23

5.  Corneal Epithelial Removal with a Newly Designed Epithelial Brush.

Authors:  Ho Seok Chung; Seung Hwan Moon; Soon-Suk Kang; Minseop Kim; Hun Lee; Hungwon Tchah; Jae Yong Kim
Journal:  J Ophthalmol       Date:  2021-11-16       Impact factor: 1.909

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.