Inappropriate discharge by an implantable cardioverter defibrillator: recognition of myopotential sensing using telemetered intracardiac electrograms.

M J Sandler1, S P Kutalek.   

Abstract

Inappropriate therapy from implantable anti-tachyarrhythmia devices is a common problem with a variety of etiologies. The verification of arrhythmias or other sensed events that precipitate defibrillating shocks is difficult with first- and second-generation devices due to the absence of sufficient data storage and the inability to examine stored and real-time intracardiac electrograms. In addition, the absence of premonitory symptoms is an unreliable marker for the appropriateness of defibrillator shocks. The incorporation of improved data storage and the ability to inspect intracardiac electrograms in newer devices have greatly increased the ability to diagnose abnormal device behavior as shown in the following case report. Inappropriate implantable cardioverter defibrillator discharge due to myopotential sensing is described. The diagnosis was facilitated by telemetered intracardiac electrograms.

Entities:  

Mesh:

Year:  1994        PMID: 7516548     DOI: 10.1111/j.1540-8159.1994.tb02401.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Inappropriate arrhythmia detection in implantable defibrillator therapy due to oversensing of diaphragmatic myopotentials.

Authors:  B Schulte; J Sperzel; J Carlsson; M Dürsch; A Erdogan; H F Pitschner; J Neuzner
Journal:  J Interv Card Electrophysiol       Date:  2001-12       Impact factor: 1.900

2.  Implantable cardioverter defibrillator-an unusual case of inappropriate discharge during showering.

Authors:  A G Manolis; A G Katsivas; C V Vassilopoulos; N E Louvros
Journal:  J Interv Card Electrophysiol       Date:  2000-04       Impact factor: 1.900

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.