Pacing system analyzers: different systems--different results.

J H Wilson1, J B Siegmund, R Johnson, S E Lattner, S P Fahner.   

Abstract

When a pacemaker is implanted, several electrophysiological parameters are measured using pacing system analyzers (PSA). Different PSAs may use different filter settings and measuring techniques when compared to the implanted pacemaker. In order to determine if there were significant differences in measurements obtained with different PSAs, we obtained measurements in a group of 99 patients with three different PSAs and a manual method. The results show that with each of the three different PSAs tested, different amplitudes of intracardiac electrograms are obtained and that they are usually higher than those obtained by manual measurement of recorded electrograms. Despite significant differences, however, all methods correlate well with each other. Following common practice of pacemaker programming, the use of a PSA for the implantation of a pacemaker that uses different sensing technique does not lead to clinical complications.

Entities:  

Mesh:

Year:  1994        PMID: 7511229     DOI: 10.1111/j.1540-8159.1994.tb01345.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Dual-chamber cardiac pacemaker tester.

Authors:  C W Galuschky; J G Webster
Journal:  Med Biol Eng Comput       Date:  1998-03       Impact factor: 2.602

Review 2.  Permanent pacemakers and the elderly: concerns, costs and benefits.

Authors:  J M McComb; R S Bexton
Journal:  Br Heart J       Date:  1995-07

3.  Investigation of the bnp level changes in blood stream in different modes and lead locations after pacemaker implementation.

Authors:  Alireza Mehri Dehnavi; Afshin Fakhrpour; Mohamad Bagher Tavakoli; Mohamad Hossein Nikoo
Journal:  J Med Signals Sens       Date:  2012-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.