BACKGROUND: Various stent models have been developed for the treatment of inoperable stenoses of the central airways caused by external compression. Increasing use is made of the silicone stents designed by Dumon. We tested their technical feasibility, tolerance, and long-term efficacy in relieving respiratory symptoms in patients referred for endoscopic palliation of malignant disease.
METHODS: All procedures were performed under general anesthesia with the use of the rigid bronchoscope. We inserted 38 stents in 31 patients (median age, 67 years; 25 men and 6 women) whose airways showed residual obstruction of > 50 percent of the lumen after laser resection of endobronchial tumor and/or mechanical dilatation of extrinsic compressions.
RESULTS: Stent placement and removal--where necessary--were easy in all patients, but five stents inserted in three patients with short (< or = 2.5 cm) and conical stenoses migrated, necessitating emergency removal. In 27 of the remaining 28 patients, stent tolerance was excellent; 1 proximal tracheal stent (< 1 cm below the vocal cords) had to be removed because of otalgia and dysphagia. One lethal hemoptysis occurred within hours after a repeated laser therapy and removal of an indwelling stent. No other serious complications occurred. Immediate and lasting relief of dyspnea and improvement in performance status (Karnofsky scale, activity index) was achieved in 90 percent (28/31) of patients (p < 0.01). The influence of adjuvant radiotherapy on local tumor recurrence and survival was analyzed in a subgroup of ten patients with stage IIIB squamous cell carcinoma with comparable performance status. Five did not undergo adjuvant radiotherapy (group A) and five did (group B). In group A, four of five stents were occluded by tumor recurrence above or below the stent after a median follow-up of 2 months; in group B, zero of five were occluded (p < 0.05) after 4 months. Median survival was 4 months in group A and 6 months in group B; the difference did not reach significance.
CONCLUSIONS: The silicone stents designed by Dumon are easily inserted and removed; they are also well tolerated and very efficacious in relieving respiratory symptoms caused by extrinsic airway compression. Short and conical stenoses present limitations for their use due to increased risk of migration. Combined treatment with laser resection, stent insertion, and subsequent radiotherapy is necessary to prevent local tumor recurrence and may improve survival.

Entities:  

Mesh:

Substances:

Year:  1993        PMID: 7504608     DOI: 10.1378/chest.104.6.1653

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  34 in total

1.  Bronchogastric fistula successfully treated with the insertion of a covered bronchial stent.

Authors:  M J Bennie; T Sabharwal; J Dussek; A Adam
Journal:  Eur Radiol       Date:  2002-10-09       Impact factor: 5.315

2.  The role of airway stent placement in the management of tracheobronchial stenosis caused by inoperable advanced lung cancer.

Authors:  Kinya Furukawa; Junzo Ishida; Gaku Yamaguchi; Jitsuo Usuda; Hidemitsu Tsutsui; Makoto Saito; Chimori Konaka; Harubumi Kato
Journal:  Surg Today       Date:  2010-03-26       Impact factor: 2.549

Review 3.  Tracheobronchial tuberculosis.

Authors:  Vikas Pathak; Ray W Shepherd; Samira Shojaee
Journal:  J Thorac Dis       Date:  2016-12       Impact factor: 2.895

Review 4.  Minimally invasive palliative interventions in advanced lung cancer.

Authors:  Christopher Mallow; Margaret Hayes; Roy Semaan; Thomas Smith; Russell Hales; Roy Brower; Lonny Yarmus
Journal:  Expert Rev Respir Med       Date:  2018-06-19       Impact factor: 3.772

Review 5.  Stents for airway strictures: selection and results.

Authors:  Adil Ayub; Adnan M Al-Ayoubi; Faiz Y Bhora
Journal:  J Thorac Dis       Date:  2017-03       Impact factor: 2.895

Review 6.  Techniques of endoscopic airway tumor treatment.

Authors:  Nicolas Guibert; Laurent Mhanna; Sylvain Droneau; Gavin Plat; Alain Didier; Julien Mazieres; Christophe Hermant
Journal:  J Thorac Dis       Date:  2016-11       Impact factor: 2.895

7.  One-lung ventilation in a patient with stenting for tracheobronchial stenosis caused by esophageal cancer.

Authors:  Masaaki Kobayashi; Ryu Okutani
Journal:  J Anesth       Date:  2011-03-02       Impact factor: 2.078

8.  Effect of metal surface characteristics on the adhesion performance of the integrated low-level energies method of adhesion.

Authors:  Toshiyuki Aodai; Toru Masuzawa; Kazuhide Ozeki; Akio Kishida; Tetsuya Higami
Journal:  J Artif Organs       Date:  2012-08-30       Impact factor: 1.731

9.  Outcome for malignant tracheobronchial stenoses in esophageal cancer.

Authors:  Shinzo Takamori; Hiromasa Fujita; Akihiro Hayashi; Masahiro Mitsuoka; Yasuhiro Terazaki; Keisuke Miwa; Mari Fukunaga; Kazuo Shirouzu
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2002-06

10.  A Dumon stent inserted for bronchial stenosis causing a left bronchopericardial fistula: report of a case.

Authors:  S Sugiyama; S Koyama; A Murakami; Y Mizushima; T Misaki; K Matsui; M Kitagawa
Journal:  Surg Today       Date:  1998       Impact factor: 2.549

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.