Comparison of transtracheal and extravascular Doppler determinations of stroke volume and cardiac output at various states of volume loading in piglets.

R J Peterson1, N Kissoon, S P Murphy, S R Goodwin, E J Bayne, E W Kelley, E L Ceithaml.   

Abstract

OBJECTIVE: To assess the applicability of a new technology in neonates. Transtracheal Doppler and extravascular Doppler determinations of stroke volume and cardiac output were compared with thermodilution measurements at various states of volume loading in an animal model.
DESIGN: Prospective, descriptive study.
SETTING: Animal research laboratory at a university medical center.
SUBJECTS: Fourteen newly weaned piglets, weighing 2.8 to 6.5 kg.
INTERVENTIONS: Doppler probes were placed on the endotracheal tube tip (transtracheal Doppler) and directly on the aortic adventitia (extravascular Doppler). A 4-Fr thermodilution catheter was inserted in the pulmonary artery. Stroke volume and cardiac output determinations were recorded at baseline, after a 15-mL/kg volume load and after successive 15-mL/kg blood withdrawals to exsanguination or a systolic blood pressure of < 20 mm Hg.
MEASUREMENTS AND MAIN RESULTS: Transtracheal and extravascular Doppler measurements of cardiac output were not significantly different from thermodilution at any physiologic state. These techniques were able to measure stroke volumes and cardiac outputs at the low levels seen in severe hemorrhagic shock.
CONCLUSIONS: Transtracheal Doppler and extravascular Doppler measurements of cardiac output compare favorably with thermodilution. These methods effectively followed trends from alterations in intravascular volume, even at very high heart rates and small stroke volumes. Transtracheal Doppler and extravascular Doppler should yield useful information in critically ill neonatal patients, where data regarding stroke volume and cardiac output may be useful in clinical management.

Entities:  

Mesh:

Year:  1995        PMID: 7497724     DOI: 10.1097/00003246-199512000-00010

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  1 in total

1.  Epi-aortic Doppler measurement of cardiac output in univentricular connection.

Authors:  Ad J J C Bogers; Martin van den Burg; Ronald Schepp; Jan Klein
Journal:  Med Devices (Auckl)       Date:  2009-02-03