Literature DB >> 7496747

Intracranial pressure monitoring techniques.

J Ghajar1.   

Abstract

In patients requiring ICP monitoring, a ventricular catheter connected to an external strain gauge transducer or catheter-tip pressure transducer device is the most accurate and reliable method of monitoring ICP, and enables therapeutic CSF drainage. Clinically significant infections or hemorrhage associated with ICP devices causing patient morbidity are rare and should not deter the decision to monitor ICP. Parenchymal catheter-tip pressure transducer devices measure ICP, similar to ventricular ICP pressure, but have the potential for significant measurement differences and drift due to the inability to recalibrate. These devices are advantageous when ventricular ICP is not obtained or if there is obstruction in the fluid coupling. Subarachnoid or subdural fluid-coupled devices and epidural ICP devices are currently less accurate.

Entities:  

Mesh:

Year:  1995        PMID: 7496747

Source DB:  PubMed          Journal:  New Horiz        ISSN: 1063-7389


  9 in total

Review 1.  Critical care issues in the early management of severe trauma.

Authors:  Alberto Garcia
Journal:  Surg Clin North Am       Date:  2006-12       Impact factor: 2.741

Review 2.  Neuromonitoring in the intensive care unit. I. Intracranial pressure and cerebral blood flow monitoring.

Authors:  Anuj Bhatia; Arun Kumar Gupta
Journal:  Intensive Care Med       Date:  2007-05-24       Impact factor: 17.440

3.  Noninvasive intracranial pressure measurement using infrasonic emissions from the tympanic membrane.

Authors:  Eduard Stettin; Klaus Paulat; Chris Schulz; Ulrich Kunz; Uwe Max Mauer
Journal:  J Clin Monit Comput       Date:  2011-08-23       Impact factor: 2.502

4.  Cerebral arterial oxygen saturation measurements using a fiber-optic pulse oximeter.

Authors:  J P Phillips; R M Langford; S H Chang; K Maney; P A Kyriacou; D P Jones
Journal:  Neurocrit Care       Date:  2010-10       Impact factor: 3.210

Review 5.  Monitoring and interpretation of intracranial pressure.

Authors:  M Czosnyka; J D Pickard
Journal:  J Neurol Neurosurg Psychiatry       Date:  2004-06       Impact factor: 10.154

Review 6.  Intrahospital transport of critically ill patients.

Authors:  C Waydhas
Journal:  Crit Care       Date:  1999-09-24       Impact factor: 9.097

7.  Regional pressure and temperature variations across the injured human brain: comparisons between paired intraparenchymal and ventricular measurements.

Authors:  Charmaine Childs; Liang Shen
Journal:  Crit Care       Date:  2015-06-23       Impact factor: 9.097

Review 8.  Advances in Intracranial Pressure Monitoring and Its Significance in Managing Traumatic Brain Injury.

Authors:  Usmah Kawoos; Richard M McCarron; Charles R Auker; Mikulas Chavko
Journal:  Int J Mol Sci       Date:  2015-12-04       Impact factor: 5.923

Review 9.  Optimal Timing of External Ventricular Drainage after Severe Traumatic Brain Injury: A Systematic Review.

Authors:  Charlene Y C Chau; Saniya Mediratta; Mikel A McKie; Barbara Gregson; Selma Tulu; Ari Ercole; Davi J F Solla; Wellingson S Paiva; Peter J Hutchinson; Angelos G Kolias
Journal:  J Clin Med       Date:  2020-06-25       Impact factor: 4.241

  9 in total