BACKGROUND AND
PURPOSE: We used near-infrared spectroscopy (NIRS) to monitor the cerebral oxygenation changes during CO2 reactivity tests.
METHODS: Fifty healthy volunteers were examined (age range, 19 to 68 years). The monitored parameters were as follows: transcranial Doppler (TCD) time-averaged middle cerebral artery flow velocity end-tidal CO2 (EtCO2); change in concentration of cerebral oxyhemoglobin (HbO2), deoxyhemoglobin (Hb), and total hemoglobin; mean arterial blood pressure; peripheral arterial oxygen saturation (SaO2); and extracranial tissue perfusion with the use of cutaneous laser-Doppler flowmetry. The examination protocol included both hypercapnia and hypocapnia. The cerebrovascular reactivity indexes were calculated as follows: TCD, relative change in flow velocity per 1 kPa increase in EtCO2; NIRS, absolute change in HbO2, Hb, and total hemoglobin concentration (micromoles per liter) per 1 kPa increase in EtCO2.
RESULTS: Mean middle cerebral artery flow velocity was found to be 58 cm/s at a mean baseline EtCO2 of 4.7 kPa. Mean cerebrovascular reactivities were as follows: TCD, 24%/kPa (SEM, 1.1); HbO2, 2.06 mumol/L per kilopascal (SEM, 0.08); Hb, -0.63 mumol/L per kilopascal (SEM, 0.09); and total hemoglobin concentration, 1.44 mumol/L per kilopascal (SEM, 0.1). Statistical analysis revealed significant correlation between reactivities calculated with the use of NIRS and TCD (P < .001). Although some fluctuations were observed in SaO2 and laser-Doppler flux, they were not correlated with either EtCO2 or NIRS.
CONCLUSIONS: NIRS signal changes in HbO2, Hb, and total hemoglobin concentration are very sensitive to alterations in EtCO2, which are largely independent of extracranial tissue perfusion. NIRS may be developed as an alternative method for testing cerebrovascular reactivity and may be of particular clinical importance when the ultrasound window is poor.

Entities:  

Mesh:

Year:  1995        PMID: 7491652     DOI: 10.1161/01.str.26.12.2285

Source DB:  PubMed          Journal:  Stroke        ISSN: 0039-2499            Impact factor:   7.914


  28 in total

1.  Simultaneous measurements of cerebral oxygenation changes during brain activation by near-infrared spectroscopy and functional magnetic resonance imaging in healthy young and elderly subjects.

Authors:  D Jannet Mehagnoul-Schipper; Bas F W van der Kallen; Willy N J M Colier; Marco C van der Sluijs; Leon J Th O van Erning; Henk O M Thijssen; Berend Oeseburg; Willibrord H L Hoefnagels; René W M M Jansen
Journal:  Hum Brain Mapp       Date:  2002-05       Impact factor: 5.038

2.  Risks for impaired cerebral autoregulation during cardiopulmonary bypass and postoperative stroke.

Authors:  M Ono; B Joshi; K Brady; R B Easley; Y Zheng; C Brown; W Baumgartner; C W Hogue
Journal:  Br J Anaesth       Date:  2012-06-01       Impact factor: 9.166

3.  Hemoglobin oxygen saturation as a marker of cerebral hemodynamics in carotid artery occlusion: an integrated transcranial doppler and near-infrared spectroscopy study.

Authors:  Fabrizio Vernieri; Mauro Silvestrini; Francesco Tibuzzi; Patrizio Pasqualetti; Claudia Altamura; Francesco Passarelli; Maria Matteis; Paolo Maria Rossini
Journal:  J Neurol       Date:  2006-06-19       Impact factor: 4.849

Review 4.  Near infrared spectroscopy: finding utility in malignant hemispheric stroke.

Authors:  Raphael Carandang; Derk W Krieger
Journal:  Neurocrit Care       Date:  2007       Impact factor: 3.210

Review 5.  The clinical validity of the absolute value of near infrared spectroscopy.

Authors:  Kenji Yoshitani; Yoshihiko Ohnishi
Journal:  J Anesth       Date:  2008-11-15       Impact factor: 2.078

Review 6.  Use of near-infrared spectroscopy in the adult.

Authors:  P J Kirkpatrick
Journal:  Philos Trans R Soc Lond B Biol Sci       Date:  1997-06-29       Impact factor: 6.237

7.  Influence of cerebral and muscle oxygenation on repeated-sprint ability.

Authors:  Kurt J Smith; François Billaut
Journal:  Eur J Appl Physiol       Date:  2010-03-31       Impact factor: 3.078

8.  Near-infrared spectroscopy assessed cerebral oxygenation during open abdominal aortic aneurysm repair: relation to end-tidal CO2 tension.

Authors:  H Sørensen; H B Nielsen; N H Secher
Journal:  J Clin Monit Comput       Date:  2015-07-04       Impact factor: 2.502

9.  Reproducibility of the blood flow index as noninvasive, bedside estimation of cerebral blood flow.

Authors:  Bendicht P Wagner; Susanne Gertsch; Roland A Ammann; Juerg Pfenninger
Journal:  Intensive Care Med       Date:  2003-01-21       Impact factor: 17.440

10.  Cerebral Autoregulation Monitoring with Ultrasound-Tagged Near-Infrared Spectroscopy in Cardiac Surgery Patients.

Authors:  Daijiro Hori; Charles W Hogue; Ashish Shah; Charles Brown; Karin J Neufeld; John V Conte; Joel Price; Christopher Sciortino; Laura Max; Andrew Laflam; Hideo Adachi; Duke E Cameron; Kaushik Mandal
Journal:  Anesth Analg       Date:  2015-11       Impact factor: 5.108