Precision of office-based blood glucose meters in screening for gestational diabetes.

S R Carr1, J Slocum, L Tefft, B Haydon, M Carpenter.   

Abstract

OBJECTIVE: Is the precision of currently available glucose meters adequate for gestational diabetes screening? STUDY
DESIGN: We performed a prospective cohort study of 62 gravid women and compared the precision of three glucose meters to laboratory standard technology.
RESULTS: The HemoCue coefficient of variation was sufficiently low in venous whole blood and plasma and capillary whole blood and plasma (3.0%, 1.8%, 2.8%, and 2.3%, respectively) to allow use in gestational diabetes screening. The OneTouch II coefficient of variation was sufficiently low in capillary whole blood (2.7%) to allow use in gestational diabetes screening but was too high in venous whole blood and plasma and in capillary plasma (3.6%, 7.8%, and 5.7%). The ExacTech coefficient of variation was too high in venous or capillary whole blood (5.4% and 8.1%) for use in gestational diabetes screening and does not analyze plasma samples.
CONCLUSION: The HemoCue meter in venous and capillary whole blood and plasma and the OneTouch II meter in capillary whole blood have coefficients of variation low enough for gestational diabetes screening.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7485335     DOI: 10.1016/0002-9378(95)91368-8

Source DB:  PubMed          Journal:  Am J Obstet Gynecol        ISSN: 0002-9378            Impact factor:   8.661


  2 in total

1.  Screening for gestational diabetes mellitus. A simple test may make it easier to study whether screening is worthwhile.

Authors:  E Rey
Journal:  BMJ       Date:  1999-09-25

2.  Prevalence of gestational diabetes mellitus among James Bay Cree women in northern Quebec.

Authors:  S Rodrigues; E Robinson; K Gray-Donald
Journal:  CMAJ       Date:  1999-05-04       Impact factor: 8.262

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.