T C Zhu1, B E Bjärngard, P Vadash.
Abstract
The presence of a wedge increases the fraction of "head-scattered" photons in a high-energy x-ray beam. We have compared internal and external wedges for x-ray beam energies between 6 and 25 MV by determining their SPRw, i.e., the ratio of the dose contribution from photons scattered by the wedge and photons either coming directly from the target or scattered by other structures, including the flattening filter. Marked differences were observed. First, SPRw for the thickest external wedge (60 degrees) was 1.8% at a field size of 10 x 10 cm2 and mildly dependent on the photon energy, while SPRw for internal wedges for this field size varied between 4.4% and 5.4% depending mostly on the location and size of the wedge and marginally on the photon energy. Second, the variation of SPRw with the collimator setting c x c was different for the internal and external wedges. SPRw for the internal wedge approached a limiting value at large c and could be fitted with an error function, while SPRw for the external wedge increased quadratically with c. As a result, the difference in SPRw (c) for internal and external wedges is reduced for large fields.Mesh:
Year: 1995 PMID: 7476722 DOI: 10.1118/1.597517
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071