Literature DB >> 7472918

Heating characteristics of the TRIPAS hyperthermia system for deep seated malignancy.

A Surowiec1, H I Bicher.   

Abstract

A deep heating hyperthermia device TRIPAS (a triapplicator system) consisting of three independent, dielectrically loaded horn applicators operating in phase at 300 MHz was investigated. The heating characteristics produced by this hyperthermia system were analyzed by means of thermochromic liquid-crystal cards and a modified CDRH (Center for Devices and Regulatory Health) elliptical phantom. Both homogenous and inhomogeneous phantoms were used, simulating high and low permittivity tissues (muscle and lung). These equivalent tissues were made of polyacrylamide gel. The semiquantitative heating pattern analysis showed a central heating of 1/3 of maximum heating at a depth of 10 cm in both homogenous (muscle) and heterogenous (muscle/fat) phantoms. Also more uniform temperature/SAR distributions were generated in muscle equivalent material than those in lung.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7472918     DOI: 10.1080/08327823.1995.11688269

Source DB:  PubMed          Journal:  J Microw Power Electromagn Energy        ISSN: 0832-7823            Impact factor:   1.325


  1 in total

1.  A preclinical system prototype for focused microwave thermal therapy of the breast.

Authors:  John Stang; Mark Haynes; Paul Carson; Mahta Moghaddam
Journal:  IEEE Trans Biomed Eng       Date:  2012-05-15       Impact factor: 4.538

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.