Advantages of CUSUM techniques for quality control in clinical chemistry.

R J Rowlands, D W Wilson, A B Nix, K W Kemp, K Griffiths.   

Abstract

The performance of routine analytical laboratories is assessed, inter alia, by the use of appropriate internal quality control techniques. Despite evidence that the cumulative sum technique is generally superior to Shewhart-type control charts and many others, its use has been limited by the popular misconception that it is inferior to Shewhart's in detecting large variations and/or outliers in quality control data. The application of computer simulation methods has enabled us to answer this criticism and has provided the basis for further improvements in the design of the appropriate control scheme for general use in clinical chemistry.

Mesh:

Year:  1980        PMID: 7471470     DOI: 10.1016/0009-8981(80)90346-0

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  1 in total

1.  CUSUM-Logistic Regression analysis for the rapid detection of errors in clinical laboratory test results.

Authors:  Maureen L Sampson; Verena Gounden; Hendrik E van Deventer; Alan T Remaley
Journal:  Clin Biochem       Date:  2015-10-30       Impact factor: 3.281

  1 in total