Literature DB >> 7448139

Photogrammetry experiments with a model eye.

A R Rosenthal, D G Falconer, I Pieper.   

Abstract

Digital photogrammetry was performed on stereophotographs of the optic nerve head of a modified Zeiss model eye in which optic cups of varying depths could be simulated. Experiments were undertaken to determine the impact of both photographic and ocular variables on the photogrammetric measurements of cup depth. The photogrammetric procedure tolerates refocusing, repositioning, and realignment as well as small variations in the geometric position of the camera. Progressive underestimation of cup depth was observed with increasing myopia, while progressive overestimation was noted with increasing hyperopia. High cylindrical errors at axis 90 degrees led to significant errors in cup depth estimates, while high cylindrical errors at axis 180 degrees did not materially affect the accuracy of the analysis. Finally, cup depths were seriously underestimated when the pupil diameter was less than 5.0 mm.

Entities:  

Mesh:

Year:  1980        PMID: 7448139      PMCID: PMC1042568          DOI: 10.1136/bjo.64.12.881

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  3 in total

1.  A NEW CAMERA FOR STEREOSCOPIC FUNDUS PHOTOGRAPHY.

Authors:  D D DONALDSON
Journal:  Trans Am Ophthalmol Soc       Date:  1964

2.  Digital photogrammetry of the optic nervehead.

Authors:  M S Kottler; A R Rosenthal; D G Falconer
Journal:  Invest Ophthalmol       Date:  1974-02

3.  Comparative reproducibility of the digital photogrammetric procedure utilizing three methods of stereophotography.

Authors:  A R Rosenthal; M S Kottler; D D Donaldson; D G Falconer
Journal:  Invest Ophthalmol Vis Sci       Date:  1977-01       Impact factor: 4.799

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.