Literature DB >> 7438151

An improved signal processor for the ultrasonic dimension gauge.

J B Sia, D G Pace, M W Osborne, M T Zanko, F Kovzelove.   

Abstract

The flip-flop producing a pulse proportional to the transit time of an ultrasonic burst between two piezoelectric crystals was often reset by transmission artefacts in sonomicrometers. A design improvement was made. The insertion of two monostable vibrators and one nand gate corrected the resetting caused by transmission artefacts. The new design produced a flip-flop pulse that was proportional to the transit time of the ultrasonic burst through the aorta. Ascending aortic diameter was monitored in unanaesthetised dogs for 6 to 8 weeks using the new design.

Entities:  

Mesh:

Year:  1980        PMID: 7438151     DOI: 10.1093/cvr/14.8.490

Source DB:  PubMed          Journal:  Cardiovasc Res        ISSN: 0008-6363            Impact factor:   10.787


  2 in total

1.  Eight-channel ultrasonic displacement meter for implantable miniature sensors.

Authors:  T Nakamura; K Hayashi; J Seki
Journal:  Med Biol Eng Comput       Date:  1987-05       Impact factor: 2.602

2.  Ultrasonic flowmeter with implantable miniature sensors.

Authors:  T Nakamura; K Hayashi; Y Taenaka; M Umezu; T Nakatani; H Takano
Journal:  Med Biol Eng Comput       Date:  1986-05       Impact factor: 2.602