Comparison of Doppler ultrasonography with arteriography of the carotid artery bifurcation.

R G Weaver, G Howard, W M McKinney, M R Ball, A M Jones, J F Toole.   

Abstract

Continuous-wave (CW) Doppler ultrasound imaging for prediction of arteriogrphic abnormality at the carotid bifurcation was carried out in 195 arteries of 105 patients. The Doppler method had no predictive value when compared to angiographic findings in arteries classified as 0-50% stenosis by Doppler. In 50-75% stenosis by arteriography Doppler accuracy was 52%. With stenosis of 76-99% Doppler imaging correlated with 71% reliability. When an arteriogram was compared with Doppler imaging the latter gave a "false negative" reading in 56% and "false positive" readings in 19%. We conclude that Doppler ultrasound evaluation provides important information regarding the state of the carotid bifurcation, which supplements the bedside evaluation, but it does not substitute for arteriography.

Entities:  

Mesh:

Year:  1980        PMID: 7414670     DOI: 10.1161/01.str.11.4.402

Source DB:  PubMed          Journal:  Stroke        ISSN: 0039-2499            Impact factor:   7.914


  3 in total

1.  Digital subtraction angiography in patients with cerebral ischaemic attacks and normal continuous wave Doppler studies.

Authors:  M Guidotti; G Landi; G Scotti; G Scarlato
Journal:  J Neurol Neurosurg Psychiatry       Date:  1985-01       Impact factor: 10.154

Review 2.  Clinical diagnosis by transcutaneous Doppler ultrasound.

Authors:  R K Wyse
Journal:  Postgrad Med J       Date:  1982-04       Impact factor: 2.401

Review 3.  Duplex ultrasound for diagnosing symptomatic carotid stenosis in the extracranial segments.

Authors:  Nicolle Cassola; Jose Cc Baptista-Silva; Luis Cu Nakano; Carolina Dq Flumignan; Ricardo Sesso; Vladimir Vasconcelos; Nelson Carvas Junior; Ronald Lg Flumignan
Journal:  Cochrane Database Syst Rev       Date:  2022-07-11
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.