| Literature DB >> 7408208 |
J M Burrin, R Worth, L A Ashworth, S Currie, K G Alberti.
Abstract
The measurement of glycosylated haemoglobin is now widely used as a guide to glycaemic control in patients with diabetes mellitus. Present methods for measurement of glycosylated haemoglobin are either laborious, expensive or both. We describe a precise and inexpensive method for the automated analysis of glycosylated haemoglobin based on a colorimetric technique. The within-batch coefficient of variation ranged from 2.0 to 4.8% and between-batch was less than 11.0%. At least 200 samples can be analysed per day.Entities:
Mesh:
Substances:
Year: 1980 PMID: 7408208 DOI: 10.1016/0009-8981(80)90373-3
Source DB: PubMed Journal: Clin Chim Acta ISSN: 0009-8981 Impact factor: 3.786