A comparison of four standard scintigraphic TV displays: concise communication.

A S Houston.   

Abstract

Four displays (pseudocolor mappings) available on a standard color TV are compared using two series of images (100 normal brains and 100 normal livers) with 50 computer-simulated lesions superimposed on each set. Four observers viewed the sets of images in such a way that the order of the display methods for both organs formed two orthogonal Latin squares. The observers were asked to locate and rate, on a standard scale, the most apparent area in each image, and ROC analysis was applied to the results. The "heated-object spectrum" was shown to be a useful display for brain images. It was also shown that the choice of display depends on the organ to be imaged.

Mesh:

Year:  1980        PMID: 7381572

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057